Summary of Contents for ResMed C-Series

  • Page 1 C-Series Heated Humidifier User Guide English...
  • Page 2 Manufacturer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Distributed by: ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA ResMed (UK) Ltd 96 Milton Park Abingdon Oxfordshire OX14 4RY UK See www.resmed.com for other ResMed locations worldwide.
  • Page 3: Intended Use

    CPAP (continuous positive airway pressure) therapy. The C-Series Heated Humidifier attaches to your ResMed C-Series flow generator with a simple latch mechanism. This user guide contains the information you need for the correct use of your C-Series Heated Humidifier. Intended Use The C-Series Heated Humidifier is indicated for humidification of the air delivered from the ResMed C-Series CPAP System.
  • Page 4 Do not attempt to use the system unless both cables are connected to the flow generator. 6. Use the power cord supplied with the C-Series Tango flow generator and connect it to the humidifier power cord socket. 7 . Plug the other end of the power cord into a power outlet. When the combined C-Series system is powered, the green indicator lights on the flow generator and humidifier turn on.
  • Page 5: Filling The Water Chamber

    Attaching the Water Chamber Press down the release latch on the humidifier and carefully slide the filled water chamber onto the heater plate. The release latch springs back to hold the water chamber in place. heater plate release latch Detaching the Water Chamber 1.
  • Page 6: Setting The Humidity Level

    Setting the Humidity Level The aim of humidification is to deliver moist air to your mask, but to avoid having condensation form in the air tubing. The temperature and humidity of the surrounding air will affect the setting you need. You can adjust the humidity setting at any time before or during therapy, to find the most comfortable and effective setting.
  • Page 7: Cleaning And Maintenance

    Cleaning and Maintenance WARNING Beware of electrocution. Do not immerse the flow generator, humidifier or power cord in water. Always unplug the device before cleaning and be sure that it is dry before plugging it back in. CAUTION Do not use any of the following to clean any part of the humidifier. These •...
  • Page 8: Troubleshooting

    Troubleshooting Problem Possible Cause Solution Device not working Power cord is not plugged in Check that the power cord is (no display on LCD properly. plugged firmly into the power screen and/or outlet and the switch (if indicator light not available) is on at the outlet.
  • Page 9 The water leaves scaly mineral Refer to the weekly cleaning deposits in the water deposits. instructions. chamber Use distilled water. Symbols which appear on the C-Series Heated Humidifier Attention, consult accompanying documents Type CF equipment Class II equipment Hot surface Do not touch...
  • Page 10: General Warnings And Cautions

    If you need information on these disposal systems, please contact your local waste administration. The crossed-bin symbol invites you to use these disposal systems. If you require information on collection and disposal of your ResMed device please contact your ResMed office, local distributor or go to www.resmed.com/environment. General Warnings and Cautions A warning alerts you to possible injury.
  • Page 11 This product (C-Series Heated Humidifier) should be inspected by an authorised ResMed Service Center 5 years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed.
  • Page 12: Technical Specifications

    1060 hPa (sea level) to 680 hPa 2591 m (8500 ft) IEC 60601-1 Classification Class II (double insulation), Type CF The C-Series Heated Humidifier is rated to IPX1 (The power cord is excluded from this classification) Electromagnetic Compatibility Refer to your flow generator user guide for details.
  • Page 13: Limited Warranty

    Excludes single-use devices. Note: Some models are not available in all regions. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does not...
  • Page 14 34841/1 2011-11 C-Series Tango User ROW enG Global leaders in sleep and respiratory medicine www.resmed.com...