The cover should be able to open and close freely while securely attached to the
docking station.
3
Replace the water chamber in the humidifier docking station (I-6).
S
ERVICING
With normal use of a H
A
3i, the water chamber needs to be replaced every
UMID
IRE
two years.
This product (H
A
3i) should be inspected by an authorised ResMed Service
UMID
IRE
Centre 5 years from the date of manufacture. Prior to this, the device is intended to
provide safe and reliable operation provided that it is operated and maintained in
accordance with the instructions provided by ResMed. Applicable ResMed warranty
details are provided with the device at the time of original supply. Of course, as with
all electrical devices, if any irregularity becomes apparent, you should exercise caution
and have the device inspected by an authorised ResMed Service Centre.