Teledyne 4M User Manual page 41

Area scan cameras
Table of Contents

Advertisement

Falcon2 4M, 8M, and 12M Camera User's Manual
GenICam Parameters
Feature
FPN Global
Backgrou nd Su btract
System Gain
PRN U correction requ ires a clean, w hite reference. The qu ality of this reference is im p ortant for p rop er
calibration. White p ap er is often not su fficient becau se the grain in the w hite p ap er w ill d istort the
correction. White p lastic or w hite ceram ic w ill lead to better balancing.
For best results, ensure that:
Let's go through a flat field calibration example:
Here is the factory calibration procedure for the 8M camera:
Teledyne DALSA
Sensor Control Menu Item >
Black Level Selector
Black Level Selector
Gain Selector
1.
Am bient light flicker (e.g. flu orescent lights) is su fficiently low not to affect
cam era p erform ance and calibration resu lts.
2.
The average p ixel shou ld be at least 20% below the target ou tp u t. If the target
is too close, then som e p ixels m ay not be able to reach fu ll sw ing d u e to
correction ap p lied by the cam era.
3.
When 6.25 % of p ixels from a single row w ithin the region of interest are
clip p ed , flat field correction resu lts m ay be inaccu rate.
4.
Correction resu lts are valid only for the cu rrent black offset valu es. If you
change this valu e, it is recom m end ed that you recalcu late you r coefficients.
1.
The cam era is p laced in full internal exposure and frame rate (no other
exp osu re m od e w ill allow FFC calibration).
2.
Settings su ch as fram e rate, exp osu re tim e, etc. are set as close as p ossible to
the actu al op erating cond itions. Set system gain to 1 and background
subtract to 0, as these are the d efau lts d u ring FFC calibration.
3.
Select correction active set to user flat field 1. Go to flat field correction
mode, select calibration.
4.
Clear existing coefficients by p ressing flat field calibration, clear coefficient.
5.
Place the cam era in the d ark and p ress FPN calibration, this p erform s the
FPN correction and saves the FPN coefficients to non-volatile m em ory.
6.
Calibration m od e enables both FPN and PRN U correction . Verify signal
ou tp u t is close to 0 DN .
7.
Illu m inate the sensor to 65 % satu ration and set flat field target to 80 %
satu ration.
8.
Press PRN U calibration.
9.
Save the flat field calibration.
1.
The cam era is p laced in fu ll internal, 8 tap s, 10 bits, active w ind ow (3328 x
2816, only available to factory), system gain 1, backgrou nd su btract 0, global
FPN calibrated su ch that d ark FPN is 30 DN (10 bit) 50 fp s, 1500 µs exp osu re.
This last p art is im p ortant (50 fp s, 1500 µs exp osu re) and ensu res that the
cam era is in non-concu rrent m od e. In non -concu rrent m od e, read ou t and
integration d o not overlap thu s elim inating som e resid u al artifacts.
Parameter
Digital ALL1
Digital ALL2
Digital ALL
03-032-20107-01
41