Table B–11: Warranted Characteristics — Triggering System; Table B–12: Warranted Characteristics — Probe Compensator Output; Table B–13: Warranted Characteristics — Power Requirements - Tektronix TDS 410A User Manual

Digitizing oscilloscopes
Hide thumbs Also See for TDS 410A:
Table of Contents

Advertisement

Appendix B: Specification
Table B–11: Warranted Characteristics — Triggering System
Name
Accuracy, Trigger Level or Threshold,
DC Coupled
Sensitivity, Edge-Type Trigger, DC
2
Coupled
Sensitivity, Video-Type, TV Field and TV
2
Line
Pulse Width, minimum, Events-Delay
Auxiliary Trigger Input, External Clock Input
Auxiliary Trigger, Maximum Input
Frequency
Frequency, External Clock
1
Net Offset = Offset – (Position Volts/Div). Net Offset is the voltage level at the center of the A-D converter dynamic
range. Offset Accuracy is the accuracy of this voltage level.
2
The minimum sensitivity for obtaining a stable trigger. A stable trigger results in a uniform, regular display triggered on
the selected slope. The trigger point must not switch between opposite slopes on the waveform, and the display must not
"roll" across the screen on successive acquisitions. The TRIG'D LED stays constantly lighted when the SEC/DIV setting
is 2 ms or faster but may flash when the SEC/DIV setting is 10 ms or slower.
Table B–12: Warranted Characteristics — Probe Compensator Output
Name
Output Voltage and Frequency,
Output Voltage and Frequency,
Probe Compensator
Probe Compensator
Table B–13: Warranted Characteristics — Power Requirements
Name
Source Voltage and Frequency
Power Consumption
B–14
Description
1
(2% of |Setting – Net Offset
| + 0.2 div volts/div setting + Offset Accuracy) for any
channel as trigger source and for signals having rise and fall times 20 ns
0.35 division from DC to 50 MHz, increasing to 1 division at 350 MHz (TDS 410A and
TDS 420A) or 500 MHz (TDS 460A) for any channel as trigger source
0.6 division of video sync signal
5 ns
Connector: BNC at rear panel
Input Load: equivalent to three TTL gate loads
Input Voltage (maximum): –5 VDC to +10 VDC (TTL levels recommended)
10 MHz
Duty Cycle
High and low levels must be stable for 50 ns
DC to 10 MHz
High and low levels must be stable for 50 ns
Description
Characteristic
Voltage
Frequency
Description
90 to 132 VAC
, continuous range, for 48 Hz through 62 Hz
RMS
100 to 132 VAC
, continuous range, for 48 Hz through 440 Hz
RMS
180 to 250 VAC
, continuous range, for 48 Hz through 440 Hz
RMS
240 Watts (370 VA)
TDS 410A, TDS 420A & TDS 460A User Manual
Limits
0.5 V (base-top) 5% into a 1 M
1 kHz 5%
W
load

Advertisement

Table of Contents
loading