5.11 Replacing Axle Tube

– The camber bar is always mounted to the bottom of
the wheelchair frame.
– There is no need to remove the two camber clamps
C when replacing the axle tube B.
Adjustable Camber System
Rear of Wheelchair
Front of Wheelchair
1171801-A
Camber Tube with Single Camber
Rear of Wheelchair
Front of Wheelchair
1.
Adjustable Camber System Only - Note the position of the
camber inserts A so they can be installed at the same position
in the new axle tube B.
Open the camber clamps C.
2.
3.
Remove the two rear wheels.
Adjustable Camber System Only - Remove the camber inserts A
4.
positioned in the axle tube B.
5.
Adjustable Camber System Only - Loosen, but do not remove
the two Phillips screws D that secure the toe adjustment rings
E to the axle tube B.
6.
Pull the existing axle tube B through one of the camber clamps
C.
7.
Adjustable Camber System Only - Remove toe adjustment rings
E from the existing axle tube B.
Pull the existing axle tube B through the other camber clamp C
8.
and remove from the wheelchair.
Slide the new axle tube B through one of the camber clamps C.
9.
Service
55