2
Eyepiece version
2.1
Converting the microscope
1 Eyepieces with 3D shutter
2 3D-CONTROL DSA-3
3 Line connection
4 3D illumination shutter
Fig. 2
Connections for eyepiece version
1 Fuse compartment
2 Voltage selector
3 Holding unit
4 Window
Fig. 3
Changing the voltage selector on rear of
3D-CONTROL DSA-3
18
Eyepiece version
Set up the microscope in accordance with
the operating instructions of the Axiovert
100/135/135 M
Equip the microscope with LD condenser
0.55 Ph1, Ph2, Ph3, DIC, DIC
Connect the components for 3D microscopy
to the 3D-CONTROL DSA-3 control unit: Fig.
2
Connect eyepieces with 3D shutter to the
EYEPIECE sockets
Connect 3D illumination shutter to the
ILLUMINATION SHUTTER socket
Connection to the line:
check wether the voltage set on the
selector at the rear of the 3D-
CONTROL DSA-3 unit complies with
the line voltage; it may be necessary to
change from 230 V to 115 V or vice
versa:
Fig. 3
Pull the power plug before changing the
fuses or the voltage
Use a small screwdriver to press out the
clamping piece of the fuse holder at the rear
of the 3D-CONTROL DSA-3 unit; lift the fuse
holder out of the fuse compartment
Remove the fuses
Pull voltage selector out of its holder
Turn voltage selector around 180° and insert
it back in its holder:
the new voltage is visible in the window
Insert the fuses listed under Technical Data
Insert fuse holder back in the fuse
compartment until click-stop
Axiovert 100/135/135 M
Prior to connection to the line,
B 40-022-4 e
3D Microscopy
09/98