To remove the display
Remove the front panel.
1
Disconnect the ribbon cable and the calibration cable from the
2
display.
Using the T10 TORX driver, remove the two screws that hold the
3
display to the deck.
Make sure that when you reinstall these screws that you use the
correct parts. If longer screws are used, they can short the system
board to ground.
As you lift the display, rotate it off the two tabs on the side of the
4
deck.
To remove the system board
Using the T10 TORX driver, remove the eight screws that hold the
1
system board to the deck (two of the screws are in the attenuator
covers).
Remove the two screws from the rear-panel interface connector and
2
the nut from the rear-panel BNC.
Disconnect the three ribbon cables and the calibration cable.
3
As you remove the system board, rotate the system board so that the
4
BNCs clear the front panel.
Replacing Parts in the Oscilloscope
Service
4–53