HP 54600B User's And Service Manual page 103

Hide thumbs Also See for HP 54600B:
Table of Contents

Advertisement

Service
Verifying Oscilloscope Performance
Press
4
Wait a few seconds for the measurement to settle (averaging is complete),
then note the Vp-p reading from the bottom of the display.
Vp-p = _______ mV.
Set the calibration factor percent of the power meter to the 1 MHz
5
value from the calibration chart on the probe, then press dB (REF)
on the power meter to set a 0 dB reference.
Change the frequency of the signal generator to the value shown
6
below for your instrument.
Table 4-8
Signal Generator Frequency Setting
Selected
Channel
Channel 1
Channel 2
Channel 3
Channel 4
*1mv/div to 5 mv/div = 100 MHz.
Set the calibration factor of the power meter to 100 MHz percent
7
value from the chart on the probe.
Adjust the amplitude of the signal generator for a power reading as close as
possible to 0.0 dB (REL). Power meter reading = ______ dB.
4–14
, then press the
Voltage
HP 54600B
HP 54601B
100 MHz
100 MHz
100 MHz
100 MHz
100 MHz
100 MHz
Vp-p
softkey.
HP 54602B *
150 MHz
150 MHz
250 MHz
250 MHz
HP 54603B
60 MHz
60 MHz

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 54601bHp 54602bHp 54603b

Table of Contents