HP 54600B User's And Service Manual page 110

Hide thumbs Also See for HP 54600B:
Table of Contents

Advertisement

Verify the external trigger sensitivity at 100 MHz (60 MHz for
External Trig
6
HP 54603B) and 100 mV p-p (HP 54600B and HP 54603B only).
Sensitivity
a
b
c
Verify the external trigger sensitivity at 25 MHz and 50 mV p-p
7
(HP 54600B and HP 54603B only).
a
b
, then press the Ext softkey.
Press
Source
Use the power splitter to connect the signal generator to the channel
1 input and to the external trigger input.
Change the output of the signal generator to 100 MHz and 200 mV p-p.
The power splitter divides the 200 mV so that 100 mV is applied to
each of the oscilloscope inputs.
The oscilloscope triggering should be stable. If the triggering is not
stable, try adjusting the trigger level. If adjusting the trigger level
makes the triggering stable, the test still passes. If adjusting the
trigger does not help, see "Troubleshooting the Oscilloscope," on
page 4–40.
Change the output of the signal generator to 25 MHz at about 100 mV.
Press
.
Autoscale
The oscilloscope triggering should be stable. If the triggering is not
stable, try adjusting the trigger level. If adjusting the trigger level
makes the triggering stable, the test still passes. If adjusting the
trigger does not help, see "Troubleshooting the Oscilloscope," on
page 4–40.
Verifying Oscilloscope Performance
Service
4–21

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 54601bHp 54602bHp 54603b

Table of Contents