Table 4-10
Equipment Required
Equipment
Signal generator
Power splitter
(HP 54600B/03B only)
Cable
Adapter
Feedthrough
Internal Trig
Press
1
Sensitivity
Connect the signal generator to channel 1.
2
Verify the trigger sensitivity at 25 MHz and 0.35 divisions.
3
a
b
c
Verify the trigger sensitivity at 1 division for the frequencies shown
4
below.
a
Critical specifications
sine waves:
25 MHz and 100 MHz–HP 54600B/HP 54601B
25 MHz and 250 MHz–HP 54602B
25 MHz and 60 MHz–HP 54603B
Outputs differ <0.15 dB
BNC, Qty 3
N (m) to BNC (f), Qty 3
50 , BNC (m) and (f)
, then press the
Setup
Set the signal generator to 25 MHz and about 50 mV.
Press
.
Autoscale
Decrease the output of the signal generator until there is 0.35 vertical
divisions of the signal displayed.
The trigger should be stable. If the triggering is not stable, try
adjusting the trigger level. If adjusting the trigger level makes the
triggering stable, the test still passes. If adjusting the trigger does not
help, see "Troubleshooting the Oscilloscope," on page 4–40.
Change the output of the signal generator to the frequency shown
below for your instrument and set amplitude to about 100 mV.
Verifying Oscilloscope Performance
Recommended
Model/Part
Tek SG 503/Tek TM 501
HP 11667B
HP 10503A
HP 1250-0780
HP 10100C
Default Setup
softkey.
Service
4–19