Change the time base to 5 ns/div.
8
Wait a few seconds for the measurement to settle (averaging is complete),
then note the Vp-p reading from the bottom of the display.
Vp-p = ______ mV.
Calculate the response using the following formula.
9
20 log
Correct the result from step 9 with any difference in the power meter
10
reading from step 7. Make sure you observe all number signs.
For example
Result from step 9 = 2.3 dB
Power meter reading from step 7 = 0.2 dB (REL)
True response = ( 2.3) ( 0.2) = 2.1 dB
The true response should be
If the result is not
page 4–40.
Repeat steps 1 to 10 for channel 2 (channels 2 to 4 on the HP 54601B
11
and HP 54602B).
When you measure the bandwidth on channels 3 and 4, use the 0.1 V/Div
range.
step 8 result
10
step 4 result
3 dB, see "Troubleshooting the Oscilloscope," on
Verifying Oscilloscope Performance
3 dB.
Service
4–15