Chapter 1 Getting Started
Making Measurements
To Set Limits of Measurements
To demonstrate how Math and Limits work together, use the Scale (10)
and Offset (1 Mega) values selected in the previous procedure "To Select
Scale and Offset." Enable Math by performing the following steps.
Scale &
Offset
Run
The result of this Math operation is a measurement of 101 MHz.
(Measurement
Scale) + Offset = Result
(10 MHz
10) + 1Mega = 101 Mega
Now, set the upper limit to 102 Mega and the lower limit to 100 Mega by
performing the following procedures. (Figure 1-1 and Figure 1-2 illustrate
the limits settings.)
Operating Guide
Display 1, 7/13/92
Display 1, 7/13/92
Freq
Ch 2
M