Example Procedure For Running The Self Test; To Configure The Rs-232 Serial Port For Printing - HP 53131A Operating Manual

225 mhz universal counter
Hide thumbs Also See for 53131A:
Table of Contents

Advertisement

Chapter 2 Operating Your Universal Counter
Using the Utility Menu

Example Procedure for Running the Self Test

1
Press and hold Recall (Utility) key, then cycle POWER key.
2
Press Recall (Utility) key until TEST: is displayed.
3
Press any one of the arrow keys to select the desired menu choice in the
menu item (for example, the ALL? menu choice).
4
Press Enter key to start the test.
5
If you are asked to do something, press Enter key when you have
completed the requested task.
A pass or fail message will be displayed.

To Configure the RS-232 Serial Port for Printing

Software pacing, baud rate, parity, data bits, and stop bits parameters for
each port are user-selectable and independent of configuration of the
other.
The table below lists the configuration factory-default values for the
RS-232 port.
Parameter
Software Pacing
Baud Rate
Parity
Data Bits
Stop Bits
Full Duplex
DTR
1
When parity is EVEN or ODD, data bits are fixed at 7. When parity is NONE, data bits are fixed at 8.
Operating Guide
Default
Possible Choices
XON
XON or NONE
9600
300, 1200, 2400, 9600, or 19200
NONE
EVEN or ODD
8
FIXED at 7
1
FIXED
ON
ON or OFF
HIGH
PACE, HIGH, LIMIT
1
NONE
FIXED at 8
2-61

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53132a

Table of Contents