The Curve Algorithm Screen; Checking The Linearization - Teledyne 2020 Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

4 Operation

4.8.2 The Curve Algorithm Screen

The Curve Algorithm is a linearization method. It provides from 1 to 9
intermediate points between the ZERO and SPAN values, which can be
normalized during calibration, to ensure a straight-line input/output transfer
function through the analyzer. (Before setting the alogorithm curve, each
range must be Zeroed and Spanned).
Each range is linearized individually, as necessary, since each range
will usually have a totally different linearization requirement.
To linearize the ranges, you must first perform the four steps indicated
at the beginning of section 4.8 Programming. You will then be in the MAIN
MENU and select ALOGORITHM.

4.8.2.1 Checking the linearization

From the MAIN MENU screen, select
Use the UP/DOWN key to select the range: 01, 02, or 03. Then press
Enter .
UP/DOWN to select and Enter
tion has been accomplished satisfactorily.
The leftmost digit (under
monitored. Use the UP/DOWN keys to select the successive points.
The
value is the input to the linearizer. It is the simulated output
of the analyzer. You do not need to actually flow gas.
The
value is the output of the linearizer. It should be the
ACTUAL concentration of the span gas being simulated.
If the
corrected. ESCAPE to return to the previous screen. Select and Enter
to Calibration Mode screen.
4-26
) is the number of the data point being
value shown is not correct, the linearization must be
Teledyne Analytical Instruments
, and Enter .
to check whether the lineariza-
Model 2020