2-8. Accessing the Fuse

To prevent instrument damage, verify that the correct fuse is
installed for the line voltage setting. Acceptable fuse types are
Bussman type MTH or Littelfuse type 312 or equivalent, of the
appropriate amperage rating. Use only a 4A, 250V fast-blow
fuse when the line voltage selection switches are set in the
range 200-240V.
A spare 4A, 250V fuse is shipped with the 5725A to ensure that one is
available for use on the 200-240V range.
The line power fuse is accessible on the rear panel. The fuse rating label to the right of
the fuse holder (labeled FUSE) shows the correct replacement fuse rating for each line
voltage setting. To check or replace the fuse, refer to Figure 2-4 and proceed as follows:
1. Turn off the POWER switch and disconnect the line power cord from ac line power.
2. Insert the blade of a standard screwdriver in the slot of the fuse holder labeled FUSE.
3. Turn the screwdriver counterclockwise until the cap and fuse pop free.
4. To reinstall the fuse, reverse this procedure.
2-8
WARNING
CAUTION
GROUNDING
FOR FIRE PROTECTION
CONNECTOR IN POWER CORD
REPLACE ONLY WITH A 250V FUSE
MUST BE CONNECTED TO
OF INDICATED RATING.
ENSURE PROTECTION FROM
ELECTRONIC SHOCK.
VOLTAGE
47-63 Hz
FUSE
SELECTION
750VA MAX
S2 S3 S4
!
!
100V
FUSE
110V
F8A 250V
(FAST)
115V
120V
S2 S3 S4
!
200V
220V
F4A 250V
230V
(FAST)
240V
CHASSIS
GROUND
S2
S3
S4
Figure 2-4. Accessing the Fuse
Caution
Note
JOHN FLUKE MFG. CO., INC.
MADE IN U.S.A.
PATENTS PENDING
NO INTERNAL USER SERVICEABLE
PARTS REFER SERVICE TO
QUALIFIED SERVICE PERSONNEL
TO CLEAN FILTER
REMOVE FROM INSTRUMENT
AND FLUSH WITH WARM
SOAPY WATER
CURRENT
OUTPUT
HI
11A MAX
20V PK
LO
MAX
5700A CALIBRATOR
CALIBRATION
ENABLE NORMAL
aq5f.eps