A Series Ii Resistance Specifications: 99 % And 95 % Confidence Levels - Fluke 5725A Instruction Manual

Fluke amplifier instruction manual
Hide thumbs Also See for 5725A:
Table of Contents

Advertisement

Table 1-10. 5700A Series II Resistance Specifications: 99 % and 95 % Confidence Levels
5700A
99%
99 % Confidence Level
Nominal
Absolute Uncertainty of Characterized Value
Value
5 °C from calibration temperature [Note 1]
24 Hours
0
50
1
85
1.9
85
10
26
19
24
100
15
190
15
1 k
11
1.9 k
11
10 k
9
19 k
9
100 k
11
190 k
11
1 M
16
1.9 M
17
10 M
33
19 M
43
100 M
110
5700A
95%
95 % Confidence Level
Nominal
Absolute Uncertainty of Characterized Value
Value
5 °C from calibration temperature [Note 1]
24 Hours
0
50
1
70
1.9
70
10
21
19
20
100
13
190
13
1 k
9
1.9 k
9
10 k
7.5
19 k
7.5
100 k
9
190 k
9
1 M
13
1.9 M
14
10 M
27
19 M
35
100 M
90
Note: 1. Specifications apply to displayed value. 4-wire connections, except 100 M .
90 Days
180 Days
ppm
50
50
95
100
95
100
28
30
26
28
17
18
17
18
12
13
12
13
11
12
11
12
13
14
13
14
18
20
19
21
37
40
47
50
120
125
90 Days
180 Days
ppm
50
50
80
85
80
85
23
27
22
24
14
15
14
15
10
11
10
11
9.5
10.5
9.5
10.5
11
12
11
12
15
17
16
18
31
34
39
42
100
105
Introduction and Specifications
1 Year
24 Hours
50
50
110
32
110
25
33
5
31
4
20
2
20
2
15
2
15
2
14
2
14
2
16
2
16
2
23
2.5
24
3.5
46
10
55
20
130
50
1 Year
24 Hours
50
50
95
32
95
25
28
5
27
4
17
2
17
2
13
2
13
2
12
2
12
2
14
2
14
2
20
2.5
21
3
40
10
47
20
110
50
Resistance Specifications
Relative Uncertainty
1 °C
90 Days
ppm
50
40
33
8
7
4
4
3.5
3.5
3.5
3.5
3.5
3.5
5
6
14
24
60
Relative Uncertainty
1 °C
90 Days
ppm
50
40
33
8
7
4
4
3.5
3.5
3.5
3.5
3.5
3.5
5
6
14
24
60
1
1-17

Hide quick links:

Advertisement

Table of Contents
loading