Table 1-2. 5700A Series II DC Voltage Specifications: 99 % and 95 % Confidence Levels
5700A
99%
99 % Confidence Level
Range
220 mV
2.2 V
11 V
22 V
220 V
1100 V
5700A
95%
95 % Confidence Level
Range
220 mV
10 nV
2.2 V
100 nV
11 V
1 mV
22 V
1 mV
220 V
10 mV
1100 V
100 mV
1-8
5 °C from calibration temperature
Resolution
For fields strengths >1 V/m but 3 V/m,
24 Hours
10 nV
6.5 + .75
100 nV
3.5 + 1.2
1 V
3.5 + 3
1 V
3.5 + 6
10 V
05 + 100
100 V
07 + 600
5 °C from calibration temperature
Resolution
For fields strengths >1 V/m but 3 V/m,
24 Hours
5.5 + 0.6
3.5 + 1
03 + 3.5
03 + 6.5
04 + 80
06 + 500
Absolute Uncertainty
add 0.01 % of range.
90 Days
180 Days
(ppm output + V)
7 + .75
08 + .75
6 + 1.2
07 + 1.2
5 + 4
07 + 4
5 + 8
07 + 8
6 + 100
08 + 100
8 + 600
10 + 600
Absolute Uncertainty
add 0.01 % of range.
90 Days
180 Days
1 Year
(ppm output + V)
6 + 0.6
7 + 0.6
8 + 0.6
5 + 1
6 + 1
7 + 1
4 + 3.5
6 + 3.5
7 + 3.5
4 + 6.5
6 + 6.5
7 + 6.5
5 + 80
7 + 80
8 + 80
7 + 500
8 + 500
9 + 500
Relative Uncertainty
1 Year
24 Hours
(ppm output + V)
09 + .8
2.5 + .5
08 + 1.2
2.5 + 1.2
08 + 4
1.5 + 3
08 + 8
1.5 + 6
09 + 100
2.5 + 100
11 + 600
03 + 600
Relative Uncertainty
24 Hours
(ppm output + V)
02 + 0.4
02 + 1
1.2 + 3
1.2 + 6
02 + 80
2.4 + 500
1 °C
90 Days
04 + .5
04 + 1.2
3.5 + 4
3.5 + 8
04 + 100
4.5 + 600
1 °C
90 Days
3.5 + 0.4
3.5 + 1
03 + 3.5
03 + 7
3.5 + 80
04 + 500

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents