Program Function Mode - Fluke Hart Scientific 9150 User Manual

Portable furnace
Hide thumbs Also See for Hart Scientific 9150:
Table of Contents

Advertisement

it can be adjusted.
00200
Press "SET" to save the new soak-time value or "EXIT" to discard changes
S
The next value to edit is the program scan rate. This value is ignored if scan is
not enabled for the unit (See Section
Sr 4
Press "SET" to edit the program scan rate.
S
10.0
Use "UP" and "DOWN" to adjust the program scan rate.
11.3
Press "SET" to save the new scan rate value or "EXIT" to discard changes.
S
After "SET" is pressed the controller advances to the next program point or, if
there are no more program points to edit, exits to the Program Function Menu.
Repeat the above steps to edit any program point.
8.4.4

Program Function Mode

The next parameter is the program function or cycle mode. There are four pos-
sible modes which determine whether the program scans up (from set-point 1
to n) only or both up and down (from set-point n to 1), and also whether the
program stops after one cycle or repeats the cycle indefinitely. The table below
shows the action of each of the four program mode settings.
Pf=1
Program point 4 soak time set for 200 minutes
Accept the program point soak time
Program point 4 scan rate
Edit the program point scan rate
Current program point 4 scan rate
New program point 4 scan rate
Accept the program point scan rate
Function
1
2
3
4
8.3.1).
Action
up-stop
up-down-stop
up-repeat
up-down-repeat
8 Controller Operation
Ramp and Soak Program
27

Advertisement

Table of Contents
loading

Table of Contents