9150 Portable Furnace
User's Guide
8.11.4
8.12
34
via the serial interface are immediately echoed or transmitted back to the device
of origin. With half duplex the commands are executed but not echoed. The du-
plex mode parameter is indicated by,
dUPL
Serial duplex mode parameter
Press "SET" to access the mode setting
Current duplex mode setting
d=FULL
The mode may be changed using "UP" or DOWN" and pressing "SET".
New duplex mode setting
d=HALF

Linefeed

The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
Serial linefeed parameter
LF
Press "SET" to access the linefeed parameter.
LF=On
Current linefeed setting
The mode may be changed using "UP" or "DOWN" and pressing "SET".
New linefeed setting
LF=OFF

Calibration Parameters

The operator of the 9150 has access to the furnace calibration constants. These
values are set at the factory and must not be altered. The correct values are im-
portant to the accuracy and proper and safe operation of the furnace. Access to
these parameters is available to the user only so that in the event the controller
memory fails, the user may restore these values to the factory settings. The user
should have a list of these constants and their settings with the manual.
CAUTION: DO NOT change the values of the furnace calibration con-
stants from the factory set values. The correct settings of these parameters
is important to the safety and proper operation of the furnace.
The calibration parameters menu is indicated by,