Major Internal Components - Teledyne 3000MA Operating Instructions Manual

Percent oxygen analyzer
Hide thumbs Also See for 3000MA:
Table of Contents

Advertisement

Percent Oxygen Analyzer
5.4
The Paramagnetic cell is accessed by unlatching and swinging open the
front panel, as described earlier. Other internal components are accessed by
removing the rear panel and sliding out the entire chassis. See Figure 5-4,
below. The gas piping is illustrated in Figure 2-4, and the major electronic
components locations are shown in Figure 2-5, in chapter 2.
WARNINGS:See warnings on the title page of this manual.
The 3000MA contains the following major components:
Analysis Section
2 x Power Supplies
Microprocessor
Displays
RS-232 Communications Port.
See the drawings in the Drawings section in back of this manual
for details.
X
X
To detach the rear panel, remove only the eight screws marked with an X.
2
Amplifier output doesn't match test input
3
Both Failed

Major Internal Components

Paramagnetic Sensor
Sample System
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
Figure 5-3: Rear-Panel Screws
Teledyne Analytical Instruments
Maintenance 5
X
X
5-3