The RS-232 protocol allows some flexibility in its implementation.
Table 3-2 lists certain RS-232 values that are required by the 3000MA
implementation.

3.3.3 Remote Probe Connector

The 3000MA is a single-chassis instrument, which has no Remote
Probe Unit. Instead, the Remote Probe connector is used as another method
for controlling external sample/zero/span gas valves. See Figure 3-6.
Due to power supply limits, the maximum combined current that can be
pulled from these output lines is 100 mA. (If two lines are ON at the same
time, each must be limited to 50 mA, etc.) If more current and/or a different
voltage is required, use a relay, power amplifier, or other matching circuitry
to provide the actual driving current.
In addition, each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This could limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-7.
Toggling input. Stops/Starts any status message output from
the RS-232, until st<enter> is sent again.
Table 3-2: Required RS-232 Options
Parameter
Baud
Byte
Parity
Stop Bits
Message Interval
Figure 3-6: Remote Probe Connector Pinouts
Teledyne Analytical Instruments
Setting
2400
8 bits
none
1
2 seconds
Installation 3