Most 326 series analyzers contain no moving parts (although some
contain an analog meter movement, solenoid, or relay contacts), and there-
fore, periodic service is not required. The sample flow check and calibration
outlined in Section 4 should be adequate to keep the analyzer functioning
throughout the life span of a given Micro-Fuel Cell.
Special designs may require sample systems with filter elements, valves,
etc., and those components will require inspection or replacement as needed
to keep the instrument in service. A periodic inspection schedule (once per
week, once per month, etc.) will be sufficient to determine problems before
they become serious.
When the analyzer is equipped with a Class A-3 cell, it must be pro-
tected from prolonged exposure (more than 4 hours) to CO
spheres (such as air). This can be accomplished by trapping CO
sample (flue gas) by closing the input manifold valves and sealing the vent
connection with a valve, water trap or clamped-off plastic tubing.
When the Micro-Fuel Cell nears the end of its useful life, sensitivity will
decline very rapidly. The initial response to this phenomenon is to
recalibrate the analyzer. If many turns of span control are required, or if the
control does not have sufficient range to recalibrate the analyzer, a new
Micro-Fuel Cell is required.
Maintenance &
Maintenance & T T T T T r r r r r oub
Maintenance &
Maintenance &
Maintenance &
5.1 Maintenance
5.1 Maintenance
5.1 Maintenance
5.1 Maintenance
5.1 Maintenance

5.1.1 Routine Maintenance

5.1.1 Routine Maintenance
5.1.1 Routine Maintenance
5.1.1 Routine Maintenance
5.1.1 Routine Maintenance
5.1.2 Special Shut-Down Precaution for Class A-3 Cell
5.1.2 Special Shut-Down Precaution for Class A-3 Cell
5.1.2 Special Shut-Down Precaution for Class A-3 Cell
5.1.2 Special Shut-Down Precaution for Class A-3 Cell
5.1.2 Special Shut-Down Precaution for Class A-3 Cell
5.1.3 Cell Replacement
5.1.3 Cell Replacement
5.1.3 Cell Replacement
5.1.3 Cell Replacement
5.1.3 Cell Replacement
Teledyne Analytical Instruments
Model 326/327/328
Model 326/327/328
Model 326/327/328
Model 326/327/328
Model 326/327/328
oub
oub
leshooting
leshooting
oubleshooting
oub
leshooting
leshooting
-free atmo-
2
-containing
2
5-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

327326

Table of Contents