NOTE:
NOTE: Never expose the sensor to varying temperatures (i.e., hold the sensor in your
NOTE:
Never expose the sensor to varying temperatures (i.e., hold the sensor in your
Never expose the sensor to varying temperatures (i.e., hold the sensor in your
Never expose the sensor to varying temperatures (i.e., hold the sensor in your
NOTE:
NOTE:
Never expose the sensor to varying temperatures (i.e., hold the sensor in your
hand) while calibrating.
hand) while calibrating.
hand) while calibrating.
hand) while calibrating.
hand) while calibrating.
1. Switch the unit ON using "0/I" ON/OFF key.
2. Place the sensor in 100% oxygen.
NOTE:
NOTE:
NOTE: The flow diverter should only be used with flowing gas circuits.
The flow diverter should only be used with flowing gas circuits.
The flow diverter should only be used with flowing gas circuits.
The flow diverter should only be used with flowing gas circuits.
NOTE:
NOTE:
The flow diverter should only be used with flowing gas circuits.
3. The CALIBRATE button is located on the top of the unit. Turn the button
by pressing down on it and twisting. Rotate until the display reads 100%.
4. Remove the sensor from 100% oxygen. If the flow diverter is attached,
remove it and expose the sensor to room air. The LCD should display 20–
22%.
5. A calibration check may be conducted at any time to assure proper opera-
tion. Remove the diverter and expose the sensor to room air, and verify that
the reading is 20–22%. Expose the sensor to 100% oxygen (preferably using
the diverter and tee adapter in a flow circuit) and verify that the reading is
99–101%.
NOTE:
NOTE:
Never calibrate the unit in humidified gas, as water vapor dilutes the calibration
Never calibrate the unit in humidified gas, as water vapor dilutes the calibration
NOTE:
NOTE: Never calibrate the unit in humidified gas, as water vapor dilutes the calibration
NOTE:
Never calibrate the unit in humidified gas, as water vapor dilutes the calibration
Never calibrate the unit in humidified gas, as water vapor dilutes the calibration
gas and makes the oxygen concentration appear lower than it really is.
gas and makes the oxygen concentration appear lower than it really is.
gas and makes the oxygen concentration appear lower than it really is.
gas and makes the oxygen concentration appear lower than it really is.
gas and makes the oxygen concentration appear lower than it really is.

Use

Use
Use
Use
Use
NOTE:
NOTE:
NOTE: Prior to use, check the calibration and check the sensor for leaks or damage.
Prior to use, check the calibration and check the sensor for leaks or damage.
Prior to use, check the calibration and check the sensor for leaks or damage.
Prior to use, check the calibration and check the sensor for leaks or damage.
NOTE:
NOTE:
Prior to use, check the calibration and check the sensor for leaks or damage.
The TED 60T can be used to analyze gas mixtures for oxygen in two basic modes:
1.) In breathing circuits or other instances where gases are flowing in tubing
circuits.
2.) In confined volumes such as incubators or tents.
When measuring for oxygen in breathing circuits, the diverter must
diverter should be screwed onto the threaded front end of the T-7 sensor. A tee adapter
Teledyne Analytical Instruments
Operations 2
Operations 2
Operations 2
Operations 2
Operations 2
must be used. The
must
must
must
2-3