Do's & Don'ts - Teledyne TED 200-T Instruction Manual

Portable oxygen monitor
Table of Contents

Advertisement

2.3
Do's and Don'ts
Read all of the directions before using for the first time.
Calibrate every 8 hours or before every use.
Visually inspect the sensor for leakage before each use.
Use 2-point calibration every time: 21%, then 100%.
Check the HI and LO alarm settings prior to each use.
Test batteries regularly (replace immediately when indicated).
Keep the unit, sensor and connections dry, or on the dry side of the breathing
circuit.
Recalibrate after replacing the batteries.
Recalibrate after replacing the sensor.
Use properly installed alkaline batteries only.
Make sure the T-7 sensor is properly attached.
Use the plastic flow diverter only when using the tee adapter.
Remove and save the plastic flow diverter when using the unit in chamber
applications (incubators, tents, etc.)
Perform an alarm test before each use (press the ALARM TEST key twice).
Clean the case with isopropyl alcohol or mild detergent only.
Use the TED 200-T if you suspect any malfunction.
Use the TED 200-T in the presence of flammable gases.
Use anything but alkaline batteries.
Autoclave or freeze the T-7 sensor or TED 200-T unit.
Open or try to repair a leaking or broken sensor.
Immerse the unit or sensor in any liquid.
Pass hot or cold gas mixtures over the sensor.
Expose the unit to devices that produce high levels of radio, short wave,
microwave, x-ray, or high frequency interference.
Use cleaning agents or liquids in the cable receptacles or around the battery
compartment.
Place the TED 200-T unit itself in a water vapor-saturated environment.
Expose the LCD to excessive sunlight.
Expose the TED 200-T to a condensing water environment such as a mist
tent.
Teledyne Analytical Instruments
– DO –
– DON'T –
Operations 2
2-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents