Any UTR Alarm Status Prefix datum is delivered as a two-byte binary bit map (16-bit, big
endian) with each bit representing the status of UTR channel #16 through UTR channel #1
respectively. A bit value of 0 (zero) indicates the UTR is operating within its specified limits. A
value of 1 (one) indicates the UTR is outside its specified limits (i.e., "in alarm"). To be "in alarm"
the channel's UTR datum value has to deviate from the average of the UTR data values of all
the TC channels, by an amount greater than a specified set point (in ûC).
Page 58
Selecting too many data groups will reduce module performance.
(hex) data selected for inclusion in each stream
packet
(reserved for Valve Status Prefix (9046 does not use)
Enable UTR Alarm Status Prefix (see alarm bit map
below)
Enable Primary Transducer (Temperature/Resistance)
EU Data (default if "05" never executed after "00")
Enable Primary Transducer (Temperature/Resistance)
A/D Counts
Enable Primary Transducer (Temperature/Resistance)
Voltages
Enable UTR Temperature/Other EU Data (ûC)
Enable UTR Temperature/Other A/D Counts
Enable UTR Temperature/Other Voltages
Model 9046 User's Manual