Table 24. Environmental Characteristics
Characteristic
Temperature
Operating
Non-operating (storage)
Humidity
Altitude
Operating
Non-operating (storage)
Vibration
Operating
Non-operating
This meter is sensitive to strong external magnetic fields.
Do not use the meter within 1-inch (2.54 cm) of strong
external magnetic fields such as that produced by the
Fluke ToolPak hanging magnet accessory. Incorrect
readings may result. If erratic behavior is noted, turn meter
off to reset.
True RMS Digital Multimeters
–10 to +50 C
–40 to +60 C
–40 to +35 C: < 80%
+35 to +40 C: < 70%
+40 to +60 C: < 55%
2,000m (6,562 ft)
For altitudes from 2,000 m up to
3,000 m (9,843 ft) derate voltage input to
600 VAC CAT III.
12,300 m (40,354 ft)
2.66 gRMS, 5 to 500 Hz, 3 axes
(10 minutes each)
3.48 gRMS, 5 to 500 Hz, 3 axes
(10 minutes each)