Accuracy Specifications; Input Characteristics; Min Max Recording Accuracy And Response Time - Fluke 113 Calibration Information Manual

Electrical multimeter
Hide thumbs Also See for 113:
Table of Contents

Advertisement

113
Calibration Information

Accuracy Specifications

Function
[1,2]
k Chek
Function
[3]
s
[3]
Ω
[3]
R
[3]
E
[1]
All k Chek voltage ranges are specified from 60 counts to 100 % of range. Because inputs below 60 counts are not specified, it is possible and
normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted
together.
Crest factor of ≤3 at 4000 counts, decreasing linearly to 1.5 at full scale.
[2]
[3]
After measuring voltage, a wait time of 1 minute is required to maintain accuracy of ohms, capacitance, diode test, and continuity.

Input Characteristics

Function
Input Impedance (Nominal)
k Chek
Open Circuit Test Voltage
Ω
R

MIN MAX Recording Accuracy and Response Time

Specified accuracy of the measurement function ±40 counts in kChek for changes >500 ms in duration, ±12 counts in Ohms for changes
>325 ms in duration. Typical 100 ms response to 80 %. Response time not specified for Capacitance.
4
Range
6.000 V
60.00 V
600.0 V
Range
600 Ω
600.0 Ω
6.000 kΩ
60.00 kΩ
2.000 V
1000 nF
10.00 μF
100.0 μF
9999 μF
~3 kΩ <300 pF
<2.7 V dc
<2.7 V dc
Resolution
DC, 45 to 500 Hz
0.001 V
0.01 V
2.0 % + 3
0.1 V
Resolution
Beeper on <20 Ω, off >250 Ω; detects opens or
1 Ω
shorts of 500 μs or longer.
0.1 Ω
0.001 kΩ
0.01 kΩ
0.001 V
1 nF
0.01 μF
0.1 μF
1 μF
Common Mode Rejection Ratio
(1 kΩ Unbalanced)
>60 dB at dc, 50 or 60 Hz
Full Scale Voltage
<0.7 V dc
2.000 V dc
Accuracy
± ([% of Reading] + [Counts])
500 Hz to 1 kHz
4.0 % + 3
Accuracy
0.9 % + 2
0.9 % + 1
0.9 % + 1
2.0 % + 3
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 μF - 1000 μF: 1.9 % + 2
>1000 μF: 5 % + 20
Normal Mode Rejection
Short Circuit Current
<350 μA
<1.0 mA