Specifications - Fluke 175 Calibration Information Manual

True rms multimeters
Table of Contents

Advertisement

175, 177, & 179
Calibration Information

Specifications

Accuracy is specified for 1 year after calibration, at operating temperatures of 18 C to 28 C, with relative humidity
at 0 % to 95 %. Accuracy specifications take the form of:
ac-coupled, true rms responding.
Maximum voltage between
any terminal and earth ground: 1000 V dc or ac rms
Surge Protection:
W Fuse for mA inputs:
W Fuse for A input:
Display:
Altitude:
Temperature:
Temperature coefficient:
Electromagnetic Compatibility
(EN 61326-1:1997):
Relative Humidity:
Battery Life:
Size (H x W x L):
Weight:
Safety Compliances:
Certifications:
4
( [ % of Reading ] + [ Counts ] ). AC conversions are
8 kV peak per IEC 61010
440 mA, 1000 V FAST Fuse, minimum interrupt rating 10 kA
11 A, 1000 V FAST Fuse, minimum interrupt rating 17 kA
Digital: 6000 counts, updates 4/sec
Bar Graph: 33 segments;
Updates 40/sec
Frequency: 10,000 counts
Capacitance: 1,000 counts
Operating: 2000 m; Storage: 12,000 m
Operating: 10 °C to +50 °C;
Storage: 40 °C to +60 °C
0.1 X (specified accuracy) / °C
(< 18 °C or > 28 °C)
In an RF field of 3 V/M, accuracy = specified accuracy except in temperature:
specified accuracy
5 °C (9 °F)
Maximum, Noncondensing
90 % to 35 °C
75 % to 40 °C
45 % to 50 °C
Alkaline: 300 hrs typical
4.3 cm x 9 cm x 18.5 cm
420 g
ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Measurement
Category III, 600 V Measurement Category IV
CAT III- Equipment designed to protect against transients in equipment in fixed-
equipment installations, such as distribution panels, feeders and short branch circuits,
and lighting systems in large buildings.
CAT IV- Equipment designed to protect user against transients from the primary
supply level, such as an electricity meter or an overhead or underground utility
service.
CSA, TÜV (EN61010), P, ; (N10140), VDE, UL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

177179