Fluke 175 Calibration Information Manual page 12

True rms multimeters
Table of Contents

Advertisement

175, 177, & 179
Calibration Information
Performance Tests
When calibrating or measuring AC current, avoid the potential for errors from coupled noise, by either:
1. Shorting the VeG to the COM input
or
2. Having NO connection to the VeG input.
1. Set the UUT Rotary Switch to the position called for in Table 2 (175) or Table 3 (177/179), and for all
measurements other than current (amps), connect the calibrator to the VeG and
If testing the milliamps function, connect the calibrator to the Meter
terminals.
If testing the amps function, connect the calibrator to the Meter
2. Referring to Table 2 or Table 3, apply the indicated calibrator output voltages to the UUT for each test.
3. Verify that the UUT displays a reading that is within the limits shown in the Meter Response column.
Test
(Switch Position)
K
AC Volts
K
Hz
1
AC Volts Frequency
12
Table 2. Performance Tests 175
5500 Output
300 mV 45 Hz
5 V 500 Hz
5 V 1 kHz
50 V 45 Hz
50 V 1 kHz
300 V 45 Hz
500 V 500 Hz
500 V 1 kHz
1000 V 45 Hz
1 V 45 Hz
5 V 50 kHz
Note
10A
Lower Limit
296.7 mV ac
4.947 V ac
4.897 V ac
49.47 V ac
48.97 V ac
296.7 V ac
494.7 V ac
489.7 V ac
987 V ac
44.94 Hz
49.94 kHz
input terminals.
COM
and
input
400 mA
COM
and
input terminals.
COM
Meter Response
Upper Limit
303.3 mV ac
5.053 V ac
5.103 V ac
50.53 V ac
51.03 V ac
303.3 V ac
505.3 V ac
510.3 V ac
1013 V ac
45.06 Hz
50.06 kHz