Fluke 175 Calibration Information Manual page 18

True rms multimeters
Table of Contents

Advertisement

175, 177, & 179
Calibration Information
If the Meter has a serial number between 83911001 – 83911240 or 84060001 and above, or if the model number
flashes on the display at turn on, use the following steps.*
Function
(Switch Position)
E
(DC Millivolts)
L
(DC Volts)
C
(AC Volts)
e
(Ohms)
O
(Diode Test)
mA
(Milliamps)
A
(Amps)
* If the Meter is not connected correctly, or if the rotary switch is in the wrong position, the Meter will beep 2 times to alert the user.
18
Table 4. Calibration Steps (continued)
Calibration Step
C-01
C-02
C-03
C-04
C-05
C-06
C-07
C-08
C-09
C-10
C-11
C-12
C-13
C-14
C-15
C-16
C-17
Input Value
600.0 mV dc
120.0 mV dc
6.000 V dc
60.00 V dc
600.0 V dc
600.0 mV, 60 Hz
600.0 V, 60 Hz
600.0
6.000 k
60.00 k
600.0 k
6.000 M
5.000 V dc
400.0 mA dc
400.0 mA ac, 60 Hz
6.000 A dc
6.000 A ac, 60 Hz

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

177179