Calibration Adjustment; Performance Tests - Fluke 175 Calibration Information Manual

True rms multimeters
Table of Contents

Advertisement

Test
F
AC Milliamps
J
AC Amps
I
m
1
DC Milliamps
I
1
DC Amps
1.
Press the YELLOW button to select this function.
2.
Does not include test lead resistance.
3.
Be sure to use correct cable.
4.
To ensure accuracy, the Meter and thermocouple adapter must be at the same temperature. Allow 20 minutes before performing the
temperature procedures.

Calibration Adjustment

Perform the calibration adjustment procedures if the Meter fails the performance test.
The Meter buttons behave as follows when the calibration mode is enabled:
Press and hold this button to test the present function. This measurement is uncalibrated and
H
may be inaccurate. This is normal.
Press and hold this button to display the required input.
M
Press this button to skip the present calibration step without accepting the input signal.
R
If a calibration step is skipped, the remaining calibration procedure may be invalidated.
This feature is not available for units with serial numbers between 83911001 – 83911240
and 84060001 and above.
Press this YELLOW button to store the calibration value and advance to the next step. This
B
button is also used to exit calibration mode after the calibration adjustment sequence is
complete.
Calibration Adjustment Procedure
Use the following steps to adjust the Meter's calibration:
For Meters with a serial number of 84060000 and below, switch the Meter to volts
Meters with a serial number of 83911001 – 83911240 and 84060001 and above, switch the Meter to
E
DC.
1. Turn the Meter over and find the Calibration Seal located near the top of the Meter (refer to Figure 4).
Table 3. Performance Tests 177/179 (continued)
5500 Output
3 mA 45 Hz
50 mA 1 kHz
400 mA 1 kHz
4.0 A 45 Hz
9.0 A 1 kHz
3 mA, 0 Hz
50 mA
-400 mA
4.0 A
-9.0 A
Meter Response
Lower Limit
2.92 mA ac
49.22 mA ac
393.7 mA ac
3.937 A ac
8.83 A ac
2.94 mA dc
49.47 mA dc
-404.3 mA dc
3.957 A dc
-9.12 A dc
Note
Note
True RMS Multimeters
Calibration Adjustment
Upper Limit
3.08 mA ac
50.78 mA ac
406.3 mA ac
4.063 A ac
9.17 A ac
3.06 mA dc
50.53 mA dc
-395.7 mA dc
4.043 A dc
-8.88 A dc
C
(
). For
AC
15