Ac Voltage Frequency Verification; Wave Generator Amplitude Verification: 1 M Output Impedance - Fluke 5522A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5522A:
Table of Contents

Advertisement

Table 9-3. SC600 Option AC Voltage Amplitude Verification (cont.)
Nominal Value
Frequency (Hz)
(V p-p)
-0.11
1000
0.5
1000
-0.5
1000
2.2
1000
-2.2
1000
11
1000
-11
1000
130
1000
-130
1000
6.599 (50 )
1000

AC Voltage Frequency Verification

Nominal Value
Frequency (Hz)
(V p-p)
2.1
10
2.1
100
2.1
1000
2.1
10000

Wave Generator Amplitude Verification: 1 M Output Impedance

Table 9-5. SC600 Option Wave Generator Amplitude Verification (1 M output impedance)
Nominal Value
Wave Shape
(V p-p)
square
0.0018
square
0.0119
square
0.0219
square
0.022
square
0.056
square
0.0899
square
0.09
square
0.155
(1 M
output impedance unless noted)
Measured Value
(V p-p)
Table 9-4. AC Voltage Frequency Verification
(1 M
output impedance unless noted)
Measured Value
(Hz)
Frequency
(Hz)
Value (V p-p)
1000
1000
1000
1000
1000
1000
1000
1000
SC600 Oscilloscope Calibration Option
Deviation (V p-p)
Deviation
(Hz)
Measured
Deviation
(V p-p)
Verification Tables
1-year Spec.
(V p-p)
0.00015
0.00054
0.00054
0.00224
0.00224
0.01104
0.01104
0.13004
0.13004
0.0165375
1-year Spec. (Hz)
0.000025
0.00025
0.0025
0.025
1-Year Spec.
(V p-p)
0.000154
0.000457
0.000757
0.00076
0.00178
0.002797
0.0028
0.00475
9
9-35

Advertisement

Table of Contents
loading

Table of Contents