Meter AC Voltage Accuracy and Frequency Response Test

To prevent possible electrical shock, fire, or personal injury, ensure that the
calibrator is in standby mode before making any connection between the
calibrator and the Test Tool. Dangerous voltages are present on the calibration
source and connection cables during these steps.
To test the ac voltage measurement accuracy:
1. Connect the Test Tool to the 5502A. See
2. Select the Test Tool setup.
M
3. Press
.
1
4. Press
E
5. Press
.
J
6. Press
to select MANUAL ranging. Use
7. Set the range to the first test point in
8. Set the 5502A to source the appropriate ac voltage.
9. Observe the reading and check that it is within the range shown under the appropriate
column.
10. Continue through the test points.
When you are finished, set the 5502A to 0 (zero) Volt and Standby.
Table 12. Meter Volts AC Measurement Verification Points
Range
500.0 mV
5.000 V
50.00 V
500.0 V
1100 V (1.1 kV)
30
XW Warning
to open the Measurement menu and select V ac.
5502A Output V AC
500.0 mV
5.000 V
50.00 V
500.0 V
1000 V
Figure
10.
j
k
and
to select the ranges.
Table
12.
Frequency
60 Hz
1 kHz
3 kHz
60 Hz
1 kHz
3 kHz
60 Hz
1 kHz
3 kHz
60 Hz
1 kHz
3 kHz
60 Hz
1 kHz
3 kHz
Meter Reading
494.0 to 506.0
486.0 to 514.0
>350.0
4.940 to 5.060
4.860 to 5.140
>3.500
49.40 to 50.60
48.60 to 51.40
>35.00
494.0 to 506.0
486.0 to 514.0
>350.0
0.980 to 1.020
0.960 to 1.040
> 0.700