Emerson World Class 3000 Instruction Bulletin page 66

Oxygen analyzer with hps heater power supply field module (for use with existing signal conditioning electronics)
Hide thumbs Also See for World Class 3000:
Table of Contents

Advertisement

Instruction Manual
Appendix B Rev. 2.2
January 1997
SYMPTOM
HEATER OVERHEATS.
NOTE: ON INITIAL STARTUP THE
CHECK IF THE YELLOW WIRE IS CONNECTED
TO POSITIVE AND THE RED TO NEGATIVE
ON BOTH "FROM ELECTRONICS" AND "FROM
PROBE" TERMINAL BLOCKS. (MODELS 218
AND 225 ONLY)
CHECK "FROM ELECTRONIC", ANALOG
HEATER TERMINALS IF THERE IS A
PULSATING NOMINAL 115 VAC.
CHECK "FROM PROBE", TERMINALS AT
PROBE HEATER IF THERE IS A PULSATING
NOMINAL 44 VOLTS.
CHECK PROBE ELECTRONICS SETPOINT
HAS BEEN CHANGED ACCORDING TO
APPLICABLE ELECTRONICS PACKAGE IB.
CHECK IF PROBE HEATER VOLTAGE
SELECTION JUMPER JM7 INSTALLED
(REMOVE JM8).
*SIMPSON MODEL 260 OR EQUIVALENT MULTIMETER.
B-6
Appendices
TEMPERATURE OF THE PROBE MAY
OVERHEAT TO A NOMINAL TEMP OF
800°C.
SET METER* ON 250 VAC SCALE.
YES
SET METER* ON 50 VAC SCALE.
YES
YES
Figure B-5. HPS Troubleshooting Flowchart, #2
CONNECT THE WIRING
NO
ACCORDING TO THE
DECALS ON THE TERMINAL
COVERS.
IF THE VOLTAGE IS
NO
CONSTANTLY ON, THEN
THE TRIAC IN THE PROBE
ELECTRONICS IS SHORTED.
IF THE VOLTAGE IS
NO
CONSTANTLY ON, THEN
THE TRIAC MODULE IN
THE HPS IS SHORTED.
Rosemount Analytical Inc.
A Division of Emerson Process Management
World Class 3000
35730003