Instruction Manual
Appendix A Rev. 3.6
July 1998
Probe lengths, nominal ................................................................. 18 inches (457 mm), 3 feet (0.91 m), 6 feet
Temperature limits in process
measurement area............................................................ 50° to 1300°F (10° to 704°C)
Standard/current output ................................................................. 4-20 mA dc signal (factory set)
O
indication (Digital display
2
and analog output) ............................................................ 0.1% O
System speed of response ............................................................ less than 3 seconds (amplifier output)
Resolution sensitivity...................................................................... 0.01% O
HPS 3000 housing ......................................................................... NEMA 4X (IP56)
Probe reference air flow................................................................. 2 scfh (56.6 L/hr) clean, dry, instrument quality
Calibration gas mixtures ................................................................ Rosemount Hagan Calibration Gas Kit Part No.
Calibration gas flow........................................................................ 5 scfh (141.6 L/hr)
HPS 3000 Power supply .............................................................. 100/110/220 ±10% Vac at 50/60 Hz
HPS 3000 Power requirement ..................................................... 200 VA
HPS 3000 Ambient Operating Temperature................................ 32° to 120°F (0° to 50°C)
Ambient operating temperature (Probe Junction Box) .................. 300°F (150°C) max
Approximate shipping weights:
18 inch (457 mm) package ............................................... 55 pounds (24.97 kg)
3 foot (0.91 m) package .................................................... 60 pounds (27.24 kg)
6 foot (1.83 m) package .................................................... 65 pounds (29.51 kg)
9 foot (2.74 m) package .................................................... 72 pounds (32.66 kg)
12 foot (3.66 m) package .................................................. 78 pounds (35.38 kg)
1
All static performance characteristics are with operating variables constant.
2
Equipment ordered utilizing this document as reference will be supplied to the USA standard design. Custom-
ers requiring the EEC standard design should request the EEC documentation and utilize its ordering data.
Temperatures over 1000°F (537°C) may affect the ease of field cell replaceability.
A-2
Appendices
Table A-1. Specifications for Oxygen Analyzing Equipment.
(1.83 m), 9 feet (2.74 m), or 12 feet (3.66 m),
depending on duct dimensions
or ±3% of reading, whichever is
2
greater using Rosemount calibration gases
transmitted signal
2
air (20.95% O
6296A27G01 contains 0.4% O
8% O
N
Nominal
2
2
Rosemount Analytical Inc.
A Division of Emerson Process Management
World Class 3000
1, 2
), regulated to 5 psi (34 kPa)
2
N
Nominal and
2
2