Model 132 Digital Electronics Setup; Figure 3-4. Main Pcb (Model 132) Eprom Replacement - Emerson World Class 3000 Instruction Bulletin

Oxygen analyzer with hps heater power supply field module (for use with existing signal conditioning electronics)
Hide thumbs Also See for World Class 3000:
Table of Contents

Advertisement

Instruction Bulletin
106-300NE Rev. 3.4
May 2000
4. Press "NUM" pushbutton.
5. Using "INC", "DEC" buttons, display
parameter 125.
6. Press "VAL" button.
7. Using "INC", "DEC" buttons, change
parameter 125 value to 15.4.
8. Press "ENT" to save new value.
3-5
MODEL 132 DIGITAL ELECTRONICS
SETUP
Before beginning operation, it is important that
the probe heater set point of the existing elec-
tronics be changed to support the World Class
3000 Probe. To convert the Model 132 Digital
Electronics Package for use with the World
Class 3000 Probe and Heater Power Supply, an
EPROM change is necessary. The replacement
3-4
Setup

Figure 3-4. Main PCB (Model 132) EPROM Replacement

EPROM needed is as identified below:
United States
United Kingdom
Germany
France
Italy
To replace the EPROM, proceed as follows:
a. Shut off and lock out power to the
electronics package.
b. Open electronics enclosure.
c. On the Main PCB, Figure 3-4, locate and
remove old EPROM.
d. Replace with new EPROM.
e. Close electronics enclosure and power up
system.
Rosemount Analytical Inc.
A Division of Emerson Process Management
World Class 3000
IM03222G01

Advertisement

Table of Contents
loading

Table of Contents