FUNCTIONAL DESCRIPTION AND
FUNCTIONAL DESCRIPTION AND
FUNCTIONAL DESCRIPTION AND
FUNCTIONAL DESCRIPTION AND
FUNCTIONAL DESCRIPTION AND
SPECIFICATIONS HI 931002
SPECIFICATIONS HI 931002
SPECIFICATIONS HI 931002
SPECIFICATIONS HI 931002
SPECIFICATIONS HI 931002
1) LCD display
2) Mode selection switch
3) coarse setting knob
4) Unit selection key
5) ON/OFF key
Range Active Drive
2.00 to 19.99 mA -1.50 to 14.00 pH
Passive Drive
2.00 to 19.99 mA -1.50 to 14.00 pH
Active Measure
0.00 to 19.99 mA -3.50 to 14.00 pH
Passive Measure
0.00 to 19.99 mA -3.50 to 14.00 pH
Resolution
0.01mA
Accuracy (@20°C/68°F)
±0.01mA
EMC Typical Deviation
Input resistance
Fuse
Battery Type/Life
9V alkaline/1600 hours of continuous use
Ext. Power Supply
Environment
0 to 50°C (32 to 122°F); 95% RH
Dimensions
180 x 83 x 40 mm (7.1 x 3.3 x 1.6")
Weight
4
6) fine setting knob
7) Ext. power adapter socket
8) "A" connector
9) "P" connector
10) "Com." connector.
0.01 pH
±0.01 pH
±1% f.s.
20
5x20 mm, 200mA, 250V
12 to 32 VDC
320 g (11.3 oz.)

OPERATIONAL GUIDE

OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
INITIAL PREPARATION
Each meter is supplied complete with a 9V battery. Slide off the
battery compartment cover on the back of the meter (see page 9),
install the battery while paying attention to its polarity.
PASSIVE DRIVE (CALIBRATOR) MODE
This mode is intended to calibrate an active receiver unit.
The HI 931002 simulates 4 to 20 mA current, powered by the active
receiver from the 4-20 mA bus. The current value displayed on the
LCD, can be adjusted through two single turn potentiometers in
increments of 0.01 mA.
• Disconnect the external power adapter from HI 931002 if present.
• Turn the meter on by pressing the ON/OFF key.
• Select the measurement unit by pressing the UNIT key.
• Switch the selection switch to the "Drive" mode.
• Wire the calibrator "Com." and "P" connectors to the Device Under
Test (D.U.T.) input contacts (see page 4, #10 and #9 respec-
tively) using the cables provided, paying attention to the polarity
("Com." to - and "P" to + terminals).
• Turn the Fine and Coarse knobs to any desired value and make
sure that the D.U.T. reflects the current value displayed on the HI
931002 LCD.
• Calibrate the D.U.T. according to manufacturer instructions if necessary.