• Insert the supplied O-ring in the membrane.
• Rinse the supplied membrane with electrolyte while shaking it gently. Refill with
clean electrolyte. Gently tap the membrane over a surface to ensure that no air
bubbles remain trapped. To avoid damaging the membrane, do not touch it with
your fingers.
• With the sensor facing down screw the cap clockwise to the end of the threads.
Some electrolyte will overflow.

1.5.2 Installation

The multisensor probe can support 3 different electrodes, DO/Temperature, EC,
pH/ORP . To make easier the installation, the 3 sensors have 3 different color
codes: white for DO, orange for EC and yellow for pH/ORP
For the correct sensors installation, procede as follow:
• Insert the sensor with the right alignment to the corresponding coloured con-
nector.
• Screw with the supplied tool the locking nut to fix
the sensor.
• After sensors are mounted it is necessary to screw
the protection sleeve for measurements or the trans-
parent beaker for calibration.
• With the meter turned off, connect the probe to
the DIN socket on the bottom of the meter by align-
ing the pins and pushing in the plug. Tighten the
nut to ensure a good connection.