Flow meter control
This flow meter kit is specially designed for use
together with the reference machine FOM 71
LAB and complies with the new EN 60456/prA1.
The kit contains all necessary tubes and
connectors for installation to a water hose. If
installation to fixed tubes is preferred be sure that
a straight tube of 150 mm before and 75 mm
behind the flow meter is mounted in order to get
a uniform flow. The tube diameter shall be the
same as the attached tubes. Be sure that the
water has no particles that can block the rotating
sensor wheel in the flow meter. Mount if
necessary a filter in the water housing in front of
the straight tube. The kit also contains a standard
battery eliminator with euroconnection and 230V.
If other connections or voltages are needed an
eliminator should be bought locally with following
specifications: Output 24 VDC stabilised, Max
330mAmp, DC connector type: 2.1 mm pin,
negative pole in centre.
• Disconnect the power supply to the machine.
• Mount the flow meter control box on the wall.
• Remove the top panel.
• Mount the flow meter on the cold water supply
Fig
pipe.
1
• Connect cable with connection X135 to the
flow meter and connect X135 on the flow
meter control box (see page 10 and 11).
• Connect the eliminator 24V (X136) to the flow
meter control box.
• If the level control A is black it shall be
Fig
exchanged to a light grey (included in the kit).
2
Part No 471 8856-02.
Remove capacitor C22.
• Make a hole in the rear panel by knocking out
the hole to the left of the water connecions.
• Connect cable 10 connection X133 to the flow
Fig
meter control box and the other end to the
2
plinth B (X22) on connections 3 and 7 via the
hole in the rear panel.
• Disconnect the quick connector A (X9) with
Fig
the wires coming from the water valves. Pull
3
cable 9 through the hole in the rear panel and
connect X9A and X9B to both sides of opened
connector X9. Connect X9C to the flow meter
control box.
Flow meter control
1
Hot water
Cold water
connection
connection
2
B
C22
3
A
1
Flow
meter
Cold water
supply pipe
A