Maintenance manual
for reference machines
Washer extactors
FOM 71 CLS, FOM 71 MP-Lab
IEC 60456 Ed 5
438 9094-21/EN
Operating manual in original language
2010.11.30

Advertisement

loading

Summary of Contents for Electrolux FOM 71 CLS

  • Page 1 Maintenance manual for reference machines Washer extactors FOM 71 CLS, FOM 71 MP-Lab IEC 60456 Ed 5 438 9094-21/EN Operating manual in original language 2010.11.30...
  • Page 3 Contents Contents Introduction .....................1:1 Wascator FOM 71 CLS Software versions ...................2:1 Installation inspection of the reference machine ........3:1 Calibration of the level control system ............4:1 Calibration of the scale system ...............5:1 Installation of laboratory sensors for temperature and ......6:1 speed connection for external weight signal Adjustment of temperature, speed and water ........7:1...
  • Page 5 The manual only deals with the approved reference machines in IEC 60456 Ed 5: Wascator FOM 71 CLS and, FOM 71 MP-Lab equipped with flow meter 472 99 0298..
  • Page 7 Wascator FOM 71 CLS...
  • Page 9 Software bug corrected. Bug does not affect testing in accordance with IEC standards. All software from 1.30 is suitable for IEC 60456 testing, but version 1.8 is recommen- ded. Contact your nearest Electrolux Laundry Service Centre for downloading instruc- tions.
  • Page 11 Installation inspection of the reference machine Wascator FOM 71 CLS Installation inspection of the reference machine Before carrying out any major test series inspect the installation and check the following terms: • The machine must be installed on a level base. The following clearances are recom- mended in order to facilitate calibration: At least 500 mm between the machine and the wall behind and min.
  • Page 12 Installation inspection of the reference machine Drain connection • Connect a 75 mm (2 1/2") flexible rubber hose to the machine’s drain pipe, ensuring a down- ward flow from the machine. Avoid sharp bends, which may prevent proper drainage. • There should be an air-gap of 50 mm between the drain hose and the drain.
  • Page 13 Calibration of the level control system Wascator FOM 71 CLS Calibration of the level control system During testing compliant with the IEC 604456 Ed5 standard, this calibration does not control the supply water values of the wash program. However, where wash programs use the scale...
  • Page 14 Calibration of level control system Calibration of level control system The level control system can be calibrated both individually and manually. Calibration starts at the bottom of the inner drum and can go up to 250 mm from that point. A ruler is supplied with the machine with which the water level above the inner drum can be read.
  • Page 15 Calibration of level control system • The ruler has two different scales, one in mm and one in RU units (exactly 2 mm/RU unit). The calibration shall be made in at least five level steps, and the first step when the water just enters the inner drum. More calibration steps can be made if requested. - Go down to ”Options menu”.
  • Page 16 Calibration of level control system SERVICE PROGRAM MAKE A CHOICE: CLEAR TRIP HOUR COUNTER CLEAR SERVICE COUNTER CLEAR WASH PROGRAM COUNTER IN CLS CLEAR WASH PROGRAM COUNTER IN SMC SCALE ADJUSTMENTS CALIB. OF LEVEL SENSOR EXIT SELECT • ”Service Mode” menu is now shown on the display.
  • Page 17 Calibration of level control system • Close the door. • Go down to ”Calibration of level sensor”. • Press ”Select”. LEVEL CALIBRATION FILL WATER TO DESIRED LEVEL LEVEL: 0 SCU OPEN • Press ”Open” to fill the machine with water until the level reaches the botton of the inner drum where the ruler is placed. • Press ”OK”. • Enter the value read on the ruler in RU units (The first setting is 0=level in drum. • Press ”OK”. • Press ”Open” to fill more water.
  • Page 18 Calibration of level control system LEVEL CALIBRATION OLD CALIBRATION WILL BE OWERWRITTEN ARE YOU SURE? PRESS OK ELSE PRESS ANY OTHER KEY • Press ”OK”. • Go down to ”Exit”. • Press ”Select”.
  • Page 19 CLS version. If a weight transducer need to be replaced, or if the scale electronics are changed, a complete calibration of the scale is required. Please consult an Electrolux Laundry System Service Centre before performing such a calibration.
  • Page 20 Calibration of the scale system Zero adjustment of the scale (weight function) The scale is factory-set and no adjustment should be necessary unless some components have to be replaced. However, a zero adjustment must be carried out as the total weight of the machine has to be programmed into the scale.

This manual is also suitable for:

Fom 71 mp-lab