M-initial

Table 4-2 Available settings for M-initial

Function position
F1
F2
F3
F4
F5
F6
F7
F8
F8A
F9
U1253B User's and Service Guide
You may select the initial measurement functions as
FACTORY or USER. The initial measurement functions and
range can be set according to
Function setting
AC V
DC V, AC V, AC+DC V
DC mV, AC mV, AC+DC mV
Ohm, nS
Diode, frequency counter
Temperature, capacitance
DC μA, AC μA, AC+DC μA
DC mA, AC mA, AC+DC mA
DC A, AC A, AC+DC A
29 different frequencies
Each rotary switch position is assigned a default
measurement function and a default measurement range.
As example, when you turn the rotary switch to the
position, the initial measurement function is diode
measurement according to the default factory setting. In
order to choose the frequency counter function you require,
press the
Changing the Default Settings
Table 4- 2
Auto or manual ranges
Auto or manual ranges
Auto or manual ranges
Auto or manual ranges
No range setting
Auto or manual ranges
Auto or manual ranges
Auto or manual ranges
Auto or manual ranges
Duty cycle = (N/256) × 100%
Pulse width = (N/256) × (1/frequency)
button.
below.
Range setting