Multi-range mp-based panel meter and controller

Advertisement

100 * 4mA/20 = 20%, therefore enter 20.00.
Example: minimum measured value = 100, that must correspond to a
retransmitted signal of 1V.
"Lo.A" (%) = 100 * ?V that in our example corresponds to:
10
100 * 1V/10 = 10%, therefore enter 10.00.
Example: maximum measured value 800, that must correspond to a
retransmitted signal of 18mA.
"Hi.A" (%) = 100 * ?mA that in our example corresponds to:
20
100 * 18mA/20 = 90%, therefore enter 90.00.
Example: maximum measured value 800, that must correspond to a
retransmitted signal of 5V.
"Hi.A" (%) = 100 * ?V that in our example corresponds to:
10
100 * 5V/10 = 50%, therefore enter 50.00.
Also in this case it is possible to invert the scale, that is, a decreasing
value of the retransmitted signal may correspond to an increasing value
of the input variable (example: Lo.A > Hi.A).
38
• S.ou: selection menu of the serial
communication output.
Add
: programmable address value:
1 < Add < 255
b.dr
: programming of the data baud
rate:
r1: 1200 bps.
r2: 2400 bps.
r3: 4800 bps.
r4: 9600 bps.
39