Glossary of displayed symbol:
18
PAS
: protection key: programmable
from 0 to 255.
• InP : input range selection menu:
r1 : from 0.001 to 500 Hz or from
2ms to 1000s;
r2 : from 0.1 to 50 KHz or from
20 s to 10s
• t.ba : time base programming menu:
values programmable from 0.1 to 999.9;
it is the updating time of both measure-
ment and display.
• oPr :operating mode selection menu:
General information:
"A" means always channel A input;
"B" means always channel B input;
the signal on the channel A is managed
always according to this formula: (Fa/
Pu.1)*PS.1; the signal on the channel
B is managed always according to this
formula: (Fa/Pu.2)*PS.2. Where: Fa or
Fb is the frequency signal measured on
channel A or B; Pu.1 or Pu.2 is the
number of pulses per revolution gener-
ated by the sensor connected to channel
A or B; PS.1 or PS.2 is the prescaler
value of the sensor A or B.
F1 : frequency, tachometer, rate mode A or B channel
Input A
HOLD
Input
Display
The instrument measures the frequency either on input A or on input B
and executes the following calculation: display=Fa/Pu.1*PS.1 (or Fb/
Pu.2*PS.2). It can be used, for example, to measure the rotation speed
of a motor, or to display frequency in Hz and kHz, as tachometer mode:
RPS, RPM, RPH, MPS, MPM, MPH and as rate meter: flow (m
mm
3
,kg/m
3
, g/cm
3
, l/s, l/min, l/h, m
(mm/s, cm/s, m/s, mm/min, cm/min, m/min, mm/h, cm/h, m/h, km/h, in/
s, ft/s, yd,s, in/min, ft/min, yd/min, in/h, ft/h, yd/h).
F2 : A - B, frequency (speed) difference
Input A - B
HOLD
Input
Display
The instrument measures the frequency of input A and input B and
executes the following calculation: display=(Fa/Pu.1*PS.1)-(Fb/
3
, cm
3
/s, m
3
/min, m
3
/h, t/h, kg/s) or speed
3
,
19