Download Print this page

DH Instruments MOLSTIC-S SUPPLY ONLY PLATFORM 1-4 IN. AND 1-2 IN Installation Sheet

Molstic-s supply only platform 1/4 in. and 1/2 in.

Advertisement

Quick Links

The 1/4 in. and 1/2 in. supply only molstic-S platforms,
P/N 401876 and 401877, provide inlet gas pressure
regulation to either the Device Under Test (DUT) or the
downstream molstic-S mounting systems. They are
used to control the pressure supplied to an upstream
DUT, or to adjust the flow rate through the system
while calibrating a downstream DUT.
High pressure gases are potentially hazardous.
Energy stored in these gases can be released
unexpectedly and with extreme force.
pressure systems should be assembled and
operated only by personnel who have been
instructed in proper safety practices.
Figure 1 1/4 in. Supply Only molstic-S
1.
Quick connect adaptor
(includes stem with 1/4 in. SWG - not shown)
2.
Pressure reducing regulator
3.
Flow metering/shut-off valve
4.
Outlet adaptor, 1/4 in. tube x 1/4 in. male VCR
molstic-S™ Supply Only Platform
High
Figure 2 1/2 in. Supply Only molstic-S
1.
2.
3.
4.
OPERATION OF SUPPLY ONLY MOLSTIC-S
Set regulator outlet pressure to minimum.
Rotate the knob of the pressure reducing regulator
in the counterclockwise direction until the spring
force of the diaphragm is relieved. Insure that the
shut-off valve is in the closed position.
Connect the gas supply.
Connect the gas supply to the quick connect stem
(either 1/4 in. or 1/2 in. tube). The gas supply
pressure should not exceed 1.7 MPa (250 psig).
Insert the stem into the quick connect body.
Connect the outlet of the supply only platform to
either the DUT or the inlet of the molstic-S.
1/4 in. and 1/2 in.
P/Ns 401876 and 401877
Installation Sheet
Quick connect adaptor
(includes stem with 1/2 in. SWG - not shown)
Pressure reducing regulator
Flow metering/shut-off valve
Outlet adaptor, 1/2 in. tube x 1/2 in. male VCR
Be sure that the molstic-S supply only
platform shut-off valve is closed prior to
connecting the quick connect stem to the
quick connect body (located on the inlet side
of the regulator).

Advertisement

loading
Need help?

Need help?

Do you have a question about the MOLSTIC-S SUPPLY ONLY PLATFORM 1-4 IN. AND 1-2 IN and is the answer not in the manual?

Questions and answers

Summary of Contents for DH Instruments MOLSTIC-S SUPPLY ONLY PLATFORM 1-4 IN. AND 1-2 IN

  • Page 1 molstic-S™ Supply Only Platform 1/4 in. and 1/2 in. P/Ns 401876 and 401877 Installation Sheet The 1/4 in. and 1/2 in. supply only molstic-S platforms, P/N 401876 and 401877, provide inlet gas pressure regulation to either the Device Under Test (DUT) or the downstream molstic-S mounting systems.
  • Page 2 The outlet pressure of the pressure reducing Align the regulator’s male VCR fitting with the regulator is controlled by adjusting the position of mating female swivel nut and thread the nut onto the control knob. Rotating the knob clockwise the fitting. raises the outlet pressure.

This manual is also suitable for:

Molstic-s401876401877