Page 2
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, Inc.
Page 5
Figure 2. Front and side views with dimensions ..................5 Figure 3. System schematic......................... 6 Figure 4. Typical PGC-10000-AF system installation .................. 8 Figure 5. Connecting GPC1-10000-AF to RPM4/HPMS A70M/A20M-AF ..........12 Figure 6. Adapting TEST port to connect a device under test..............13 Figure 7.
WONDER ABOUT ALL THE GREAT FEATURES YOU MIGHT BE MISSING, GET INTO THE MANUAL! GPC1-1000-AF is usually delivered as part of an PGC-10000-AF system which also includes an RPM4/HPMS A70M/A20M-AF reference pressure monitor and and a GB-152-AF gas booster. The RPM4/HPMS and the GB-152 have their own Operation and Maintenance Manuals.
GPC1 combines the versatility, speed and reliability of direct operator control with the convenience and effort-free operation of automation. It is the standard pressure control component in a PGC-10000-AF Pneumatic Gauge Calibrator as well as in a PG7202 piston gauge system.
GB-152-AF: Gas booster package used to supply gas pressure up to 10 000 psi (70 MPa) to the GPC1-10000-AF pressure controller. Each of the three components of the PGC-10000-AF system has its own Operation and Maintenance Manual and individual setup and start up instructions. Figure 4 shows the typical setup configuration of the complete PGC-10000-AF system.
2. INSTALLATION SITE REQUIREMENTS The GPC1-10000-AF is typically delivered as part of a PGC-10000-AF Pneumatic Gauge Calibrator (see Section 2.2). The PGC-10000-AF system also includes a GB-152-AF gas booster and an RPM4/HPMS A70M/A20M that have their own Operation and Maintenance Manuals.
GPC1 requires a high pressure gas supply of 10 000 psi (70 MPa) or lower. If the GPC1 is part of a PGC-10000-AF Pneumatic Gauge Calibrator, the high gas pressure supply is provided by a GB-152-AF gas booster. Hardware to connect the booster output to the GPC1 SUPPLY port is supplied with the booster.
2.4.4.2 CONNECTING TO A DEVICE UNDER TEST In a PGC-10000-AF system, the device under test (DUT) is intended to be connected to one of the TEST connections on the top surface of the GPC1. See Section 2.4.4.3 concerning cleanliness considerations before connecting a DUT.
2. INSTALLATION The PGC-10000-AF fittings accessory kit includes adaptors to convert the DH500 F TEST connection to other commonly used fittings. The adaptors are made by combining a DH500 F adaptor with a 2.75 in., 6 in. or 12 in. DH500 nipple. The...
• TEST2 port (left lateral): Normally, the system reference pressure measurement device is connected here. In a PGC-10000-AF Pneumatic Gauge Calibrator, this is an RPM4/HPMS A70M/A20M-AF. • TEST3 port (top, front): This port is connected to the lowest point on the GPC1 sump and should be used for connecting test devices that may contain liquids.
If the GPC1 is part of a PGC-10000-AF Pneumatic Gauge Calibrator, the RPM4/HPMS A70M/A20M-AF leak test function can be used (see the RPM4/HPMS Operation and Maintenance Manual).
GPC1 is most often used to set and adjust pressure to a reference measuring device and a DUT when performing a test or calibration. If the GPC1 is part of a PGC-10000-AF Pneumatic Gauge Calibrator, the reference measurement device is an RPM4/HPMS A70M/A20M. The typical operational sequence is as...
GPC1-10000-AF™ OPERATION AND MAINTENANCE MANUAL WARRANTY STATEMENT Except to the extent limited or otherwise provided herein, DH Instruments, Inc. (DHI) warrants for one year from purchase, each new product sold by it or one of its authorized distributors, only against defects in workmanship and/or materials under normal service and use.