R&S UPV / UPP Table of Contents Table of Contents 1 Information on the Current Version and History......3 Version 3.3.0.0724 ......................3 Version 3.2.0.683 ......................4 Version 3.1.2.0626 (UPV only)..................5 Version 3.1.0.0571 ......................7 2 Firmware Update................18 Local Firmware Update .....................19 Remote Firmware Update..................20 Firmware Installation ....................23 3 Open Source Acknowledgement .............
R&S UPV / UPP Error! Style not defined. 1 Information on the Current Version and History 1.1 Version 3.3.0.0724 Component FW version UPxCheckPreconditions 1.3.2 UPVCompass_2.1.77.0-Release 2.1.77.0 Setup_3.3.00724Release 3.3.0 New Functionality Supporting UPP-B4 HDMI / digital audio measurements In the "Start Condition" section of the analyzer config panel the selection of “Max FFT Size”...
R&S UPV / UPP Error! Style not defined. Known Issues The remote control commands MMEM:MSIS and MMEM:CDIR, which could be used to define a default folder for file storage, do not work properly. The default folder is always c:\upv\bin and cannot be changed with those commands.
R&S UPV / UPP Error! Style not defined. Protocol analysis in line with IEC 60958 now supported with software option UPP-K21. Digital Audio option UPP-B2 required. 1/n octave analysis now available with software option UPP-K601. output impedance (switchable) and 600 input impedance (switchable) is now available for instruments with serial numbers 100100 or higher.
Page 6
R&S UPV / UPP Error! Style not defined. Component FW version UPVCompass_2.1.77.0-Release 2.1.77.0 Setup_3.1.20626Release 3.1.2 New Functionality Analyzer/Generator Generator function Stereo Sine: Filter and Equalizer can be independently selected for each channel. Generator function Play: Filter can be independently selected for each channel. Analyzer function THD: If all harmonics are outside the upper bandlimit, then the THD result is now set to “invalid”.
R&S UPV / UPP Error! Style not defined. Analog analyzer: 250 kHz bandwidth: special combinations of filters together with very long measurement times might cause the measurement not to terminate. Fixed by reducing the maximum measurement times. Generator overload might sporadically occur, if UPV-B3 is installed and generator bandwidth is changed at high output signal levels.
Page 8
R&S UPV / UPP Error! Style not defined. displayed in the status bar, if averaging is selected and if the averaging takes more than only a few seconds. The progress information starts after the 2nd average step and is updating once per second.
Page 9
R&S UPV / UPP Error! Style not defined. Remote control commands: SOURce:VOLTage:SELect VLRT SOURce:VOLTage 0.25 V SOURce:VOLTage:RATio 0.5 Ch1&2 voltage setting Remote control commands: SOURce:VOLTage:SELect VLVL SOURce:VOLTage 0.15 V SOURce:VOLTage:CH2Stereo 0.30 V Generator Sweep: The new menu item Halt defines the frequency of the generator after the sweep has terminated.
Page 10
R&S UPV / UPP Error! Style not defined. It is intended for use in mobile phone tests, to play an activation signal (usually a voice-like signal stored in a *.wav file) before the test signal (e.g. a sinewave) is started. Measurement starts, with the start of the test signal, and can be further delayed with the start condition delay.
Page 11
R&S UPV / UPP Error! Style not defined. or noise. Noise density can now also be displayed in the unit "dBV/ZHz". The reference must be set to "Noise Density dB" in the FFT graph config panel. Analyzer function FFT: The RMS level via FFT can now be calculated within a freely selectable frequency range Remote control command: SENSe[1]:FREQuency:LIMit ON | OFF OFF is the same as the previously supported wideband mode (default).
Page 12
R&S UPV / UPP Error! Style not defined. The FFT triggered mode is activated in the analyzer FFT function . If not yet active, the waveform is switched on. The trigger parameters are the same as for the waveform function, and hve to be defined there.
Page 13
R&S UPV / UPP Error! Style not defined. when the generator is set to Output Type Unbal. Graphics Grafic Windows can have a title above the grid. The title is defined in the graphic config panel: The title of a graphic is stored (and loaded) with the trace file. Numeric display Up to now the channel names were fixed to Ch1, Ch2, ……..
Page 14
R&S UPV / UPP Error! Style not defined. was only supported for active graphics window. General The setup of the instrument can be stored as *.html file, by selecting “Report Files” as file type when saving the setup….. …..and loaded with a browser Trigger Input now implemented The impuls at the trigger input acts like a keystroke of one of the front panel keys START, SINGLE or STOP/CONT.
Page 15
R&S UPV / UPP Error! Style not defined. mode. Now the signal in the loaded file is displayed. Analog analyzer bandwidth 250 kHz: If a prefilter and 3 function filters (if possible for the selected measurement function) were used, the measurement did not terminate.
Page 16
R&S UPV / UPP Error! Style not defined. DISPlay:BARgraph1|2:A|B:LEGend:DESCription "String" The following remote control commands have been missing for the PESQ display: DISPlay:PESQ:X:LABel:AUTO ON | OFF DISPlay:PESQ:X:LABel:USER 'MyXLabel' DISPlay:PESQ:X:UNIT S | MIN | DS DISPlay:PESQ:X:UNIT:AUTO ON | OFF DISPlay:PESQ:X:UNIT:USER 'MyXUnit' The following remote control commands have been missing for the PESQ display: DISPlay:PESQ1|2:A|B:LABel:AUTO ON | OFF...
Page 17
R&S UPV / UPP Error! Style not defined. Known Issues The remote control commands MMEM:MSIS and MMEM:CDIR, which could be used to define a default folder for file storage, do not work properly. The default folder is always c:\upv\bin and cannot be changed with those commands.
R&S UPV / UPP Error! Style not defined. 2 Firmware Update ATTENTION For any update of UPV firmware from a version 1.x to a version higher than 2.0.0 the upgrade to version 2.0.0 is required before. For this purpose a bootable CD “UPV Audio Analyzer Firmware Upgrade 2.0.0”...
R&S UPV / UPP Error! Style not defined. 2.1 Local Firmware Update Save the current version It is recommended to save the currently installed version. This can be done very easily and completely intuitive with the built-in recovery and backup tool. Switch off the instrument and switch it on again.
R&S UPV / UPP Error! Style not defined. not switch off the instrument, otherwise your backup may be corrupt. It’s safe to switch off the instrument with the Exit and Shutdown button, when the menu is shown again. If you want to restore a previously saved version select Restore Backup in the same way.
Page 21
R&S UPV / UPP Error! Style not defined. Connect the remote instrument to your PC The remote instrument may be connected to your PC via GPIB or via a LAN connection. Install the new firmware version • Run Setup_3.3.00724Release.exe on your PC •...
Page 22
R&S UPV / UPP Error! Style not defined. perform a rescan. Please be careful and check twice if you have selected the correct ATTENTION instruments. Depending on your company’s network structure also instruments of other departments will show up! • Additonal help will be displayed after clicking the button “Help”...
R&S UPV / UPP Error! Style not defined. 2.3 Firmware Installation Connect mouse and keyboard to the instrument. Connect external monitor to UPP (DVI monitor) or UPV66 (VGA monitor). Switch on the instrument. Plug in the USB memory stick which contains the update files. Remove the previous firmware version Use the Windows tool “Add or Remove Programs”...
Page 24
R&S UPV / UPP Error! Style not defined. Install the new firmware version on the UPV / UPP Double-click the item UPVCompass_2.1.77.0-Release.exe Keep the defaults and start the installation by selecting Next > Install. Quit the installation by selecting Exit. Skip a request for a reboot, which may be displayed on screen.
R&S UPV / UPP Error! Style not defined. 3 Open Source Acknowledgement This instrument firmware makes use of valuable open source software packages. The most important of them are listed below together with their corresponding open source license. The verbatim license texts are provided on the documentation disk delivered with each instrument.
R&S UPV / UPP Error! Style not defined. 4 Customer Support Technical support – where and when you need it For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer Support Centers. A team of highly qualified engineers provides telephone support and will work with you to find a solution to your query on any aspect of the operation, programming or applications of Rohde &...