Shodex HILICpak VT-50 2D Operation Manual

Advertisement

(Please read this operation manual carefully to achieve the best and consistent column performance for a long time.)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the column package and the column surface for any damage.
(2) Please check the product name and serial number (Serial no. or S/N) written on the column package and
adhesive label on the column body.
(3) Please download the Certificate of Analysis (CoA) for the purchased product. The CoA can be downloaded
from Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number.
1. Introduction
Thank you for purchasing the Shodex product. Shodex HILICpak VT-50 2D is a polymer-based hydrophilic
chromatography (HILIC) chlromatography column modified with amino group. It is suitable for separation and
analysis of anion compounds such as phosphorylated sugars and organic acids.
2. Column Components
Please refer to the Shodex website: https://www.shodex.com/en/da/07.html
3. Column Specifications
Product
Product Name
Code
F7630400
HILICpak VT-50 2D
F6711300
HILICpak VT-50G 2A
Base Material
: Spherical porous particles of polyvinyl alcohol modified with quaternary amino group
Column Housing : PEEK
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : 25 mM Formic Acid aq. Solution/Acetonitrile =15/85
4. Usable Conditions
Product Name
HILICpak VT-50 2D
HILICpak VT-50G 2A
Usable solvents are listed below.
(1) A mixture of water and acetonitrile or water and methanol of any ratio can be used. Up to 100 % water,
acetonitrile, and methanol can be used. Use of organic solvents other than acetonitrile or methanol is not
warrantied.
Operation Manual
Shodex HILICpak VT-50 2D
Column Size (mm)
I.D.
2.0
2.0
Flow Rate (mL/min)
Recommended
Maximum
0.2 - 0.3
-
Resonac Corporation (https://www.shodex.com/)
Particle Size
(μm)
Length
150
5
10
5
Maximum
Pressure
(MPa/column)
0.5
10
-
-
ver.20221202E
Theoretical Plate Number
(Per Column)
≥ 4,500
(Guard Column)
pH
Temperature
Range
Range (°C)
2 - 13
4 - 60
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HILICpak VT-50 2D and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shodex HILICpak VT-50 2D

  • Page 1 Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number. 1. Introduction Thank you for purchasing the Shodex product. Shodex HILICpak VT-50 2D is a polymer-based hydrophilic chromatography (HILIC) chlromatography column modified with amino group. It is suitable for separation and analysis of anion compounds such as phosphorylated sugars and organic acids.
  • Page 2: Eluent Preparation

    It may lead to abnormal peak shapes, poor separations, and/or low reproducibility. In such cases, please adjust the sample concentration and/or the injection volume. Note · Use of a guard column is recommended to protect the analytical column. Resonac Corporation (https://www.shodex.com/)
  • Page 3: Column Installation

    7. Column Usage Procedure HILICpak VT-50 2D is a semi-micro column. To achieve its best performance, please use it with a semi- micro type HPLC system. 7.1 HPLC System Preparation Wash the entire HPLC system prior to column installation, including all flow-lines and sample loop by switching the valves, and then replace the washing solution with the eluent to be used.
  • Page 4: Column Cleaning

    (2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction. Set the flow rate at 0.1 mL/min. Method 1: General cleaning Washing Solution Cleaning Time 6 min 0.1 M NaOH aq. solution 75 min 12 min Eluent 40 min Resonac Corporation (https://www.shodex.com/)
  • Page 5 · Never allow inside the column to dry. It can damage the column. 9. Column Inspection Please refer to the inspection method described in the CoA. At Shodex, “half width method” is adopted for the calculation of plate count and “asymmetry factor (Fas)” is adopted for the calculation of peak symmetry. Please refer to the Shodex website for the detail: https://www.shodex.com/en/da/07.html...

This manual is also suitable for:

Hilicpak vt-50g 2aF7630400F6711300

Table of Contents