Shodex HILICpak VG-25 4D Operation Manual

Advertisement

(Please read this operation manual carefully to achieve the best and consistent column performance for a long time.)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the column package and the column surface for any damage.
(2) Please check the product name and serial number (Serial no. or S/N) written on the column package and
adhesive label on the column body.
(3) Please download the Certificate of Analysis (CoA) for the purchased product. The CoA can be downloaded
from Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number.
1. Introduction
Thank you for purchasing the Shodex product. Shodex HILICpak VG-25 series is polymer-based hydrophilic
chromatography (HILIC) chlromatography columns modified with amino group. It is suitable for separation and
analysis of sugars and sugar alcohols.
2. Column Components
Please refer to the Shodex website: https://www.shodex.com/en/da/07.html
3. Column Specifications
Product
Product Name
Code
F7631000
HILICpak VG-25 4D
F6711800
HILICpak VG-25G 4A
Base Material
: Spherical porous particles of polyvinyl alcohol modified with amino group
Column Housing : PEEK
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Water/Acetonitrile = 20/80
4. Usable Conditions
Product Name
HILICpak VG-25 4D
HILICpak VG-25G 4A
Usable solvents are listed below.
(1) A mixture of water and acetonitrile or water and methanol of any ratio can be used. Up to 100 % water,
acetonitrile, and methanol can be used. Use of organic solvents other than acetonitrile or methanol is not
warrantied.
Operation Manual
Shodex HILICpak VG-25 4D
Column Size (mm)
I.D.
4.6
4.6
Flow Rate (mL/min)
Recommended
Maximum
0.3 - 0.8
-
Resonac Corporation (https://www.shodex.com/)
Particle Size
(μm)
Length
150
2.5
10
2.5
Maximum
Pressure
(MPa/column)
1.0
15
-
-
ver.20250116E
Theoretical Plate Number
(Per Column)
≥ 16,500
(Guard Column)
pH
Temperature
Range
Range (°C)
2 - 13
4 - 60
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HILICpak VG-25 4D and is the answer not in the manual?

Questions and answers

Summary of Contents for Shodex HILICpak VG-25 4D

  • Page 1 Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number. 1. Introduction Thank you for purchasing the Shodex product. Shodex HILICpak VG-25 series is polymer-based hydrophilic chromatography (HILIC) chlromatography columns modified with amino group. It is suitable for separation and analysis of sugars and sugar alcohols.
  • Page 2: Eluent Preparation

    It may lead to abnormal peak shapes, poor separations, and/or low reproducibility. In such cases, please adjust the sample concentration and/or the injection volume. Note · Use of a guard column is recommended to protect the analytical column. Resonac Corporation (https://www.shodex.com/)
  • Page 3: Column Installation

    (new and current) eluents, and then replace it with the new solvent. e.g. When replacing from water/acetonitrile to a highly concentrated buffer solution, first run water and then replace it with the eluent. Resonac Corporation (https://www.shodex.com/)
  • Page 4: Column Cleaning

    · Strong alkaline solvents such as 0.1-M aqueous sodium hydroxide solution can damage the detector cell. Do not connect the detector while regenerating the column and collect elute directly from the column outlet to the waste container. Resonac Corporation (https://www.shodex.com/)
  • Page 5 · Never allow inside the column to dry. It can damage the column. 9. Column Inspection Please refer to the inspection method described in the CoA. At Shodex, “half width method” is adopted for the calculation of plate count and “asymmetry factor (Fas)” is adopted for the calculation of peak symmetry. Please refer to the Shodex website for the detail: https://www.shodex.com/en/da/07.html...

This manual is also suitable for:

Hilicpak vg-25g 4aF7631000F6711800

Table of Contents