Page 2
• RIGOL products are covered by P.R.C. and foreign patents, issued and pending. • RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at the company's sole decision. • Information in this publication replaces all previously released materials.
Page 6
Safety Requirement replacement especially to circuits or accessories must be performed by RIGOL authorized personnel. • Provide Adequate Ventilation. Inadequate ventilation may cause an increase of temperature in the instrument, which would cause damage to the instrument. So please keep the instrument well ventilated and inspect the air outlet and the fan regularly.
Document Overview This manual is designed to guide you to properly test the performance specifications of RIGOL DP900 Series Programmable DC Power Supply. For the operation methods mentioned in the test procedures, refer to User Guide of this product. Publication Number...
Press the front-panel power key. The instrument starts and performs the self-test. If the self-test fails, shut down the instrument and restart the instrument after 5 s. If the self-test still fails, please contact RIGOL. CAUTION Before connecting to power, make sure that the rear-panel voltage selector setting matches the input AC power source and that the current fuse matches the input voltage.
Values lists the rated output values and maximum output values of the voltage and current of each channel for various models of DP900 series. Table 3.2 DP900 Channel Voltage and Current Values Rated Output Max.