NORMAL OPERATION
Plug the pump into AC wall outlet.
BATTERY OPERATION
Unplugging the pump from the AC line will place the pump on battery power. To
recharge the battery, plug the pump into an AC wall outlet. Pump will automati-
cally begin charging.
PROGRAMMING-These general instructions are for inserting the Kangaroo
924 Feeding Sets into the Kangaroo™ 924 Feeding Pump.
1. Close roller clamp on pump set.
2. Fill feeding bag or spike container prior to setting up pump.
3.
In order to achieve proper accuracy, the fluid line in the feeding bag must be
28 inches above the top of the feeding pump when initiating the feeding cycle.
4. Slowly open control clamp and fill entire line with fluid, leaving a few drops
of fluid in the drip chamber.
NOTE: Avoid filling the drip chamber more than half full.
5. Close roller clamp.
6. Load set per arrow diagram on pump. Do not overstretch tubing.
7. Press and wait while pump performs a system check.
ON
8. Check VOL CLR light or press VOL then VOL CLR to clear volume.
9. Set feed rate using one of the two following options:
• Pressing the increase or decrease button, or
• Turning the knob on the left side of the pump
Setting the Volume to be Delivered "VTBD" Rate (optional feature)
• Press the VTBD button.
• Set "VTBD" by using knob on side of pump, or by pressing the
t
increase or decrease button.
• Only use the "VTBD" option if you want the pump to stop and alarm once a
set amount of formula is delivered.
10 . Open roller clamp.
11. P ress START/HOLD .
TO CHANGE RATE:
START
1. Press .
HOLD
2. Set new rate by pressing the or keys on the front panel.
START
3. Press .
HOLD
See pump operating manual for detailed instructions.
924
t
START
HOLD
t

Advertisement

loading

Summary of Contents for Covidien Kangaroo 924

  • Page 1 Kangaroo ™ Enteral Feeding Pump Directions For Use NORMAL OPERATION Plug the pump into AC wall outlet. BATTERY OPERATION Unplugging the pump from the AC line will place the pump on battery power. To recharge the battery, plug the pump into an AC wall outlet. Pump will automati- cally begin charging. PROGRAMMING-These general instructions are for inserting the Kangaroo ™ 924 Feeding Sets into the Kangaroo™ 924 Feeding Pump. 1. Close roller clamp on pump set. 2. Fill feeding bag or spike container prior to setting up pump. In order to achieve proper accuracy, the fluid line in the feeding bag must be 28 inches above the top of the feeding pump when initiating the feeding cycle. 4. Slowly open control clamp and fill entire line with fluid, leaving a few drops of fluid in the drip chamber.
  • Page 2: Troubleshooting Guide

    Connector or Optix lock ™ ™ window. If "dots" appear with message; Move Pump away from bright light. too much ambient light System Rotor is not turning while pump is Return pump for technical service. Error running. Alarm (SYS Err) Rotor is turning but is not detected Return pump for technical service. by rotor sensors. No Audible Battery is fully discharged. Charge battery for a minimum of 12 hours. Alarm Alarm circuitry failure. Return pump for technical service. COVIDIEN, COVIDIEN with Logo and ™ marked brands are trademarks of Covidien AG or its affiliate. ©2008 Covidien AG or its affiliate. All rights reserved. H-6201-xM-0108 15 Hampshire Street 1-800-962-9888 www.covidien.com Mansfield, MA 508-261-8000 02048...